Biomechanical Analysis of a New Carbon Fiber/Flax/Epoxy Bone Fracture Plate Shows Less Stress Shielding Compared to a Standard Clinical Metal PlateSource: Journal of Biomechanical Engineering:;2014:;volume( 136 ):;issue: 009::page 91002Author:Bagheri, Zahra S.
,
Tavakkoli Avval, Pouria
,
Bougherara, Habiba
,
Aziz, Mina S. R.
,
Schemitsch, Emil H.
,
Zdero, Radovan
DOI: 10.1115/1.4027669Publisher: The American Society of Mechanical Engineers (ASME)
Abstract: Femur fracture at the tip of a total hip replacement (THR), commonly known as Vancouver B1 fracture, is mainly treated using rigid metallic bone plates which may result in “stress shielding†leading to bone resorption and implant loosening. To minimize stress shielding, a new carbon fiber (CF)/Flax/Epoxy composite plate has been developed and biomechanically compared to a standard clinical metal plate. For fatigue tests, experiments were done using six artificial femurs cyclically loaded through the femoral head in axial compression for four stages: Stage 1 (intact), stage 2 (after THR insertion), stage 3 (after plate fixation of a simulated Vancouver B1 femoral midshaft fracture gap), and stage 4 (after fracture gap healing). For fracture fixation, one group was fitted with the new CF/Flax/Epoxy plate (n = 3), whereas another group was repaired with a standard clinical metal plate (Zimmer, Warsaw, IN) (n = 3). In addition to axial stiffness measurements, infrared thermography technique was used to capture the femur and plate surface stresses during the testing. Moreover, finite element analysis (FEA) was performed to evaluate the composite plate's axial stiffness and surface stress field. Experimental results showed that the CF/Flax/Epoxy plated femur had comparable axial stiffness (fractured = 645 آ±â€‰67 N/mm; healed = 1731 آ±â€‰109 N/mm) to the metalplated femur (fractured = 658 آ±â€‰69 N/mm; healed = 1751 آ±â€‰39 N/mm) (p = 1.00). However, the bone beneath the CF/Flax/Epoxy plate was the only area that had a significantly higher average surface stress (fractured = 2.10 آ±â€‰0.66 MPa; healed = 1.89 آ±â€‰0.39 MPa) compared to bone beneath the metal plate (fractured = 1.18 آ±â€‰0.93 MPa; healed = 0.71 آ±â€‰0.24 MPa) (p < 0.05). FEA bone surface stresses yielded peak of 13 MPa at distal epiphysis (stage 1), 16 MPa at distal epiphysis (stage 2), 85 MPa for composite and 129 MPa for metalplated femurs at the vicinity of nearest screw just proximal to fracture (stage 3), 21 MPa for composite and 24 MPa for metalplated femurs at the vicinity of screw farthest away distally from fracture (stage 4). These results confirm that the new CF/Flax/Epoxy material could be a potential candidate for bone fracture plate applications as it can simultaneously provide similar mechanical stiffness and lower stress shielding (i.e., higher bone stress) compared to a standard clinical metal bone plate.
|
Collections
Show full item record
| contributor author | Bagheri, Zahra S. | |
| contributor author | Tavakkoli Avval, Pouria | |
| contributor author | Bougherara, Habiba | |
| contributor author | Aziz, Mina S. R. | |
| contributor author | Schemitsch, Emil H. | |
| contributor author | Zdero, Radovan | |
| date accessioned | 2017-05-09T01:05:35Z | |
| date available | 2017-05-09T01:05:35Z | |
| date issued | 2014 | |
| identifier issn | 0148-0731 | |
| identifier other | bio_136_09_091002.pdf | |
| identifier uri | http://yetl.yabesh.ir/yetl/handle/yetl/154060 | |
| description abstract | Femur fracture at the tip of a total hip replacement (THR), commonly known as Vancouver B1 fracture, is mainly treated using rigid metallic bone plates which may result in “stress shielding†leading to bone resorption and implant loosening. To minimize stress shielding, a new carbon fiber (CF)/Flax/Epoxy composite plate has been developed and biomechanically compared to a standard clinical metal plate. For fatigue tests, experiments were done using six artificial femurs cyclically loaded through the femoral head in axial compression for four stages: Stage 1 (intact), stage 2 (after THR insertion), stage 3 (after plate fixation of a simulated Vancouver B1 femoral midshaft fracture gap), and stage 4 (after fracture gap healing). For fracture fixation, one group was fitted with the new CF/Flax/Epoxy plate (n = 3), whereas another group was repaired with a standard clinical metal plate (Zimmer, Warsaw, IN) (n = 3). In addition to axial stiffness measurements, infrared thermography technique was used to capture the femur and plate surface stresses during the testing. Moreover, finite element analysis (FEA) was performed to evaluate the composite plate's axial stiffness and surface stress field. Experimental results showed that the CF/Flax/Epoxy plated femur had comparable axial stiffness (fractured = 645 آ±â€‰67 N/mm; healed = 1731 آ±â€‰109 N/mm) to the metalplated femur (fractured = 658 آ±â€‰69 N/mm; healed = 1751 آ±â€‰39 N/mm) (p = 1.00). However, the bone beneath the CF/Flax/Epoxy plate was the only area that had a significantly higher average surface stress (fractured = 2.10 آ±â€‰0.66 MPa; healed = 1.89 آ±â€‰0.39 MPa) compared to bone beneath the metal plate (fractured = 1.18 آ±â€‰0.93 MPa; healed = 0.71 آ±â€‰0.24 MPa) (p < 0.05). FEA bone surface stresses yielded peak of 13 MPa at distal epiphysis (stage 1), 16 MPa at distal epiphysis (stage 2), 85 MPa for composite and 129 MPa for metalplated femurs at the vicinity of nearest screw just proximal to fracture (stage 3), 21 MPa for composite and 24 MPa for metalplated femurs at the vicinity of screw farthest away distally from fracture (stage 4). These results confirm that the new CF/Flax/Epoxy material could be a potential candidate for bone fracture plate applications as it can simultaneously provide similar mechanical stiffness and lower stress shielding (i.e., higher bone stress) compared to a standard clinical metal bone plate. | |
| publisher | The American Society of Mechanical Engineers (ASME) | |
| title | Biomechanical Analysis of a New Carbon Fiber/Flax/Epoxy Bone Fracture Plate Shows Less Stress Shielding Compared to a Standard Clinical Metal Plate | |
| type | Journal Paper | |
| journal volume | 136 | |
| journal issue | 9 | |
| journal title | Journal of Biomechanical Engineering | |
| identifier doi | 10.1115/1.4027669 | |
| journal fristpage | 91002 | |
| journal lastpage | 91002 | |
| identifier eissn | 1528-8951 | |
| tree | Journal of Biomechanical Engineering:;2014:;volume( 136 ):;issue: 009 | |
| contenttype | Fulltext |