    Time Action Analysis of Catheter Manipulation During Navigation Tasks in Bifurcations

    Source: Journal of Medical Devices:;2013:;volume( 007 ):;issue: 004::page 44501
    Author:
    Clogenson, H. C. M.
    ,
    Dankelman, J.
    ,
    van den Dobbelsteen, J. J.
    DOI: 10.1115/1.4025188
    Publisher: The American Society of Mechanical Engineers (ASME)
    Abstract: Endovascular intervention is a form of minimally invasive intervention that allows catheters to be placed in practically any location of the vascular tree. However, to provide access to all these remote locations, an extensive array of catheters is needed. A specific catheter is choose based on experience, without any objective indication of its suitability during the actual procedure (Bakker, N. H., Tanase, D., Reekers, J. A., and Grimbergen, C. A., 2002, “Evaluation of Vascular and Interventional Procedures with Time—Action Analysis: A Pilot Study,â€‌ J. Vasc. Intervent. Radiol., 13(5), pp. 483–488). The aim of this study is to evaluate several catheters using timeaction analysis during a navigation task in bifurcations of various geometries. The relation between the geometry of bifurcations, the catheters, and the time taken to perform specific actions is investigated. Nine novices manipulated five widely used selective catheters with a 0.035â€‌ guidewire in a model. In the model, four bifurcations of various diameters and angles were selected. Each bifurcation was cannulated six times with two different yet suitable catheters. The participants had no direct vision of the model but navigated the instruments using the images that were captured by a camera and displayed on a screen. All images presented to the participant were recorded and used for detailed timeaction analysis of the various actions to cannulate a branch (e.g., catheter or guidewire retracted, rotated, and advanced). On average, the participants needed 28.3 s to cannulate a branch. When the ratio between the diameter of the main and side branch was high, the average time per task increased significantly, as did the number of attempts to navigate into a branch. However, neither the choice between the two suitable catheters for each bifurcation, nor the angles of the bifurcation made a significant difference in navigation time. Timeaction analysis enabled objective measurement of the time spent on various actions to cannulate a branch. The results revealed that most time was spent on retracting and rotating the catheter. This was comparable for all catheters and branches, showing that all the instruments were manipulated in a similar way and presented the same difficulties.
    • Download: (610.8Kb)
    • Show Full MetaData Hide Full MetaData
    • Get RIS
    • Item Order
    • Go To Publisher
    • Statistics

      Time Action Analysis of Catheter Manipulation During Navigation Tasks in Bifurcations

    URI
    https://yetl.yabesh.ir/yetl1/handle/yetl/152843
    Collections
    • Journal of Medical Devices

    Show full item record

    contributor authorClogenson, H. C. M.
    contributor authorDankelman, J.
    contributor authorvan den Dobbelsteen, J. J.
    date accessioned2017-05-09T01:01:43Z
    date available2017-05-09T01:01:43Z
    date issued2013
    identifier issn1932-6181
    identifier othermed_007_04_044501.pdf
    identifier urihttp://yetl.yabesh.ir/yetl/handle/yetl/152843
    description abstractEndovascular intervention is a form of minimally invasive intervention that allows catheters to be placed in practically any location of the vascular tree. However, to provide access to all these remote locations, an extensive array of catheters is needed. A specific catheter is choose based on experience, without any objective indication of its suitability during the actual procedure (Bakker, N. H., Tanase, D., Reekers, J. A., and Grimbergen, C. A., 2002, “Evaluation of Vascular and Interventional Procedures with Time—Action Analysis: A Pilot Study,â€‌ J. Vasc. Intervent. Radiol., 13(5), pp. 483–488). The aim of this study is to evaluate several catheters using timeaction analysis during a navigation task in bifurcations of various geometries. The relation between the geometry of bifurcations, the catheters, and the time taken to perform specific actions is investigated. Nine novices manipulated five widely used selective catheters with a 0.035â€‌ guidewire in a model. In the model, four bifurcations of various diameters and angles were selected. Each bifurcation was cannulated six times with two different yet suitable catheters. The participants had no direct vision of the model but navigated the instruments using the images that were captured by a camera and displayed on a screen. All images presented to the participant were recorded and used for detailed timeaction analysis of the various actions to cannulate a branch (e.g., catheter or guidewire retracted, rotated, and advanced). On average, the participants needed 28.3 s to cannulate a branch. When the ratio between the diameter of the main and side branch was high, the average time per task increased significantly, as did the number of attempts to navigate into a branch. However, neither the choice between the two suitable catheters for each bifurcation, nor the angles of the bifurcation made a significant difference in navigation time. Timeaction analysis enabled objective measurement of the time spent on various actions to cannulate a branch. The results revealed that most time was spent on retracting and rotating the catheter. This was comparable for all catheters and branches, showing that all the instruments were manipulated in a similar way and presented the same difficulties.
    publisherThe American Society of Mechanical Engineers (ASME)
    titleTime Action Analysis of Catheter Manipulation During Navigation Tasks in Bifurcations
    typeJournal Paper
    journal volume7
    journal issue4
    journal titleJournal of Medical Devices
    identifier doi10.1115/1.4025188
    journal fristpage44501
    journal lastpage44501
    identifier eissn1932-619X
    treeJournal of Medical Devices:;2013:;volume( 007 ):;issue: 004
    contenttypeFulltext
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian
     
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian