    The Influence of the Pelvic Bone on the Computational Results of the Acetabular Component of a Total Hip Prosthesis

    Source: Journal of Biomechanical Engineering:;2010:;volume( 132 ):;issue: 005::page 54503
    Author:
    Sara Barreto
    ,
    Jacinto Monteiro
    ,
    João Folgado
    ,
    Paulo R. Fernandes
    DOI: 10.1115/1.4001031
    Publisher: The American Society of Mechanical Engineers (ASME)
    Abstract: The computational models developed to evaluate the hip joint performance usually neglect the presence of the pelvic bone. However, deformation depends on the stiffness of the underlying bone, and thus, the inclusion of the pelvic bone in the model influences the computed contact pressure and wear. This work discusses the influence of the pelvic bone, and how it depends on the acetabular component stiffness. It was modeled as two different polyethylene acetabular cups, considering or not a metal-backing for both 28 mm and 32 mm diametric cups. Two finite element models are developed, considering either the acetabular component rigidly fixed or attached to the deformable bone. Results present 28% and 42% difference on the contact pressure for a polyethylene cup without metal-backing when the support conditions are changed, for the 28 mm and 32 mm cups, respectively. Linear wear results present 21% and 31% difference for the same type of cups of 28 mm and 32 mm, correspondingly. The numerical results obtained in the present work show that to model the pelvic bone of the patient with a metal-backed cup did not greatly affect contact pressures and linear wear. However, when a total hip replacement is performed with an all-polyethylene acetabular cup, the presence of the pelvic bone in the model has a major influence.
    keyword(s): Pressure , Wear , Metals , Bone AND Hip joint prostheses ,
    • Download: (302.1Kb)
    • Show Full MetaData Hide Full MetaData
    • Get RIS
    • Item Order
    • Go To Publisher
    • Statistics

      The Influence of the Pelvic Bone on the Computational Results of the Acetabular Component of a Total Hip Prosthesis

    URI
    https://yetl.yabesh.ir/yetl1/handle/yetl/142629
    Collections
    • Journal of Biomechanical Engineering

    Show full item record

    contributor authorSara Barreto
    contributor authorJacinto Monteiro
    contributor authorJoão Folgado
    contributor authorPaulo R. Fernandes
    date accessioned2017-05-09T00:36:39Z
    date available2017-05-09T00:36:39Z
    date copyrightMay, 2010
    date issued2010
    identifier issn0148-0731
    identifier otherJBENDY-27136#054503_1.pdf
    identifier urihttp://yetl.yabesh.ir/yetl/handle/yetl/142629
    description abstractThe computational models developed to evaluate the hip joint performance usually neglect the presence of the pelvic bone. However, deformation depends on the stiffness of the underlying bone, and thus, the inclusion of the pelvic bone in the model influences the computed contact pressure and wear. This work discusses the influence of the pelvic bone, and how it depends on the acetabular component stiffness. It was modeled as two different polyethylene acetabular cups, considering or not a metal-backing for both 28 mm and 32 mm diametric cups. Two finite element models are developed, considering either the acetabular component rigidly fixed or attached to the deformable bone. Results present 28% and 42% difference on the contact pressure for a polyethylene cup without metal-backing when the support conditions are changed, for the 28 mm and 32 mm cups, respectively. Linear wear results present 21% and 31% difference for the same type of cups of 28 mm and 32 mm, correspondingly. The numerical results obtained in the present work show that to model the pelvic bone of the patient with a metal-backed cup did not greatly affect contact pressures and linear wear. However, when a total hip replacement is performed with an all-polyethylene acetabular cup, the presence of the pelvic bone in the model has a major influence.
    publisherThe American Society of Mechanical Engineers (ASME)
    titleThe Influence of the Pelvic Bone on the Computational Results of the Acetabular Component of a Total Hip Prosthesis
    typeJournal Paper
    journal volume132
    journal issue5
    journal titleJournal of Biomechanical Engineering
    identifier doi10.1115/1.4001031
    journal fristpage54503
    identifier eissn1528-8951
    keywordsPressure
    keywordsWear
    keywordsMetals
    keywordsBone AND Hip joint prostheses
    treeJournal of Biomechanical Engineering:;2010:;volume( 132 ):;issue: 005
    contenttypeFulltext
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian
     
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian