    The Biomechanical Consequence of Insufficient Femoral Component Lateralization and Exposed Cancellous Bone in Hip Resurfacing Arthroplasty

    Source: Journal of Biomechanical Engineering:;2010:;volume( 132 ):;issue: 008::page 81011
    Author:
    Michael Olsen
    ,
    Edward T. Davis
    ,
    Cari M. Whyne
    ,
    Rad Zdero
    ,
    Emil H. Schemitsch
    DOI: 10.1115/1.4001159
    Publisher: The American Society of Mechanical Engineers (ASME)
    Abstract: Insufficient lateralization of the femoral component coupled with exposed reamed cancellous bone has been speculated to predispose to femoral neck fracture. The current study examined the effect of mediolateral implant position and exposed cancellous bone on the strength of the resurfaced proximal femur. Composite femurs were prepared in three configurations: (1) partial, with the implant placed at the native femoral head offset of the femur, partially exposing reamed cancellous bone; (2) proud, with a medialized implant exposing a circumferential ring of cancellous bone; and (3) complete, with a lateralized implant covering all reamed cancellous bone. Specimens were loaded to failure in axial compression. A finite element model was used to further explore the effect of exposed cancellous bone, cement mantle thickness, and relative valgus orientation on the strain distributions in the resurfaced femur. The proud group (2063 N) was significantly weaker than both the partial (2974 N, p=0.004) and complete groups (5899 N, p=0.001) when tested to failure. The partial group was also significantly weaker than the complete group when tested to failure (p=0.001). The finite element model demonstrated increasing levels of strain in the superior reamed cortical-cancellous bone interface with increasing degree of exposed cancellous bone. The condition of the femoral component medialized as the result of a thick cement mantle had the greatest detrimental impact on strain level in the superior reamed cancellous bone while a valgus oriented implant provided a protective effect. This study provides biomechanical evidence that exposed reamed cancellous bone significantly reduces the load-to-failure and increases maximum strains in the resurfaced proximal femur. The perceived benefit of reconstructing the femur to its native geometry may inherently weaken the proximal femur and increase femoral neck fracture risk if the femoral component is not sufficiently lateralized to cover all unsupported reamed cancellous bone. Relative valgus orientation of the implant may help to minimize the risk of neck fracture if reamed cancellous bone remains exposed following implant impaction.
    keyword(s): Stress , Biomechanics , Bone , Failure , Fracture (Process) , Arthroplasty , Composite materials AND Cements (Adhesives) ,
    • Download: (474.3Kb)
    • Show Full MetaData Hide Full MetaData
    • Get RIS
    • Item Order
    • Go To Publisher
    • Statistics

      The Biomechanical Consequence of Insufficient Femoral Component Lateralization and Exposed Cancellous Bone in Hip Resurfacing Arthroplasty

    URI
    https://yetl.yabesh.ir/yetl1/handle/yetl/142573
    Collections
    • Journal of Biomechanical Engineering

    Show full item record

    contributor authorMichael Olsen
    contributor authorEdward T. Davis
    contributor authorCari M. Whyne
    contributor authorRad Zdero
    contributor authorEmil H. Schemitsch
    date accessioned2017-05-09T00:36:32Z
    date available2017-05-09T00:36:32Z
    date copyrightAugust, 2010
    date issued2010
    identifier issn0148-0731
    identifier otherJBENDY-27159#081011_1.pdf
    identifier urihttp://yetl.yabesh.ir/yetl/handle/yetl/142573
    description abstractInsufficient lateralization of the femoral component coupled with exposed reamed cancellous bone has been speculated to predispose to femoral neck fracture. The current study examined the effect of mediolateral implant position and exposed cancellous bone on the strength of the resurfaced proximal femur. Composite femurs were prepared in three configurations: (1) partial, with the implant placed at the native femoral head offset of the femur, partially exposing reamed cancellous bone; (2) proud, with a medialized implant exposing a circumferential ring of cancellous bone; and (3) complete, with a lateralized implant covering all reamed cancellous bone. Specimens were loaded to failure in axial compression. A finite element model was used to further explore the effect of exposed cancellous bone, cement mantle thickness, and relative valgus orientation on the strain distributions in the resurfaced femur. The proud group (2063 N) was significantly weaker than both the partial (2974 N, p=0.004) and complete groups (5899 N, p=0.001) when tested to failure. The partial group was also significantly weaker than the complete group when tested to failure (p=0.001). The finite element model demonstrated increasing levels of strain in the superior reamed cortical-cancellous bone interface with increasing degree of exposed cancellous bone. The condition of the femoral component medialized as the result of a thick cement mantle had the greatest detrimental impact on strain level in the superior reamed cancellous bone while a valgus oriented implant provided a protective effect. This study provides biomechanical evidence that exposed reamed cancellous bone significantly reduces the load-to-failure and increases maximum strains in the resurfaced proximal femur. The perceived benefit of reconstructing the femur to its native geometry may inherently weaken the proximal femur and increase femoral neck fracture risk if the femoral component is not sufficiently lateralized to cover all unsupported reamed cancellous bone. Relative valgus orientation of the implant may help to minimize the risk of neck fracture if reamed cancellous bone remains exposed following implant impaction.
    publisherThe American Society of Mechanical Engineers (ASME)
    titleThe Biomechanical Consequence of Insufficient Femoral Component Lateralization and Exposed Cancellous Bone in Hip Resurfacing Arthroplasty
    typeJournal Paper
    journal volume132
    journal issue8
    journal titleJournal of Biomechanical Engineering
    identifier doi10.1115/1.4001159
    journal fristpage81011
    identifier eissn1528-8951
    keywordsStress
    keywordsBiomechanics
    keywordsBone
    keywordsFailure
    keywordsFracture (Process)
    keywordsArthroplasty
    keywordsComposite materials AND Cements (Adhesives)
    treeJournal of Biomechanical Engineering:;2010:;volume( 132 ):;issue: 008
    contenttypeFulltext
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian
     
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian