    MRI-Based Characterization of Bone Anatomy in the Human Knee for Size Matching of a Medial Meniscal Implant

    Source: Journal of Biomechanical Engineering:;2010:;volume( 132 ):;issue: 010::page 101008
    Author:
    Jonathan J. Elsner
    ,
    Farshid Guilak
    ,
    Avi Shterling
    ,
    Eran Linder-Ganz
    ,
    Sigal Portnoy
    DOI: 10.1115/1.4002490
    Publisher: The American Society of Mechanical Engineers (ASME)
    Abstract: Allograft or synthetic menisci have been suggested as a means to restore contact pressures following meniscectomy. However, when the natural meniscus is severely damaged/absent, the necessary size cannot be determined according to the recipient size and there is a need to estimate it from magnetic resonance imaging (MRI) of the contralateral knee or the injured knee bones. The use of the contralateral-knee for size matching is problematic due to economic and practical reasons. Hence, there are significant advantages for a sizing algorithm based only on the candidate knee geometry. The aim of this study is to characterize midrange values and variability of knee dimensions and to develop a set of mathematical relations representing knee dimensions using a minimum of imaging-based bone measurements. Tibia, femur, and meniscus measurements were taken in 118 MRI scans and used to develop a representative parametric knee model in which all dimensions are expressed using tibia plateau width. The model was verified by comparing the predicted values to direct MRI measurements for 20 additional subjects by means of the Pearson correlation and and (1986, “Statistical Methods for Assessing Agreement Between Two Methods of Clinical Measurement,” Lancet, 1, pp. 307–310) plot. Anatomical parameters in the male knee were significantly larger (∼17%) compared with corresponding female measurements. However, most relations between tibia, femur, and meniscus measurements (43/56) were not significantly different between male and female populations (p≥0.05), indicating that differences between male and female joints are generally related to scaling and not shape. Dimensions predicted by the knee model were in a good agreement with dimensions measured directly from the MRI (R2>0.96) and the Bland and Altman plot indicated that ∼95% of data points were well within the ±2 standard deviation lines of agreement. The model proposed in this study is advantageous in being able to describe typical knee proportions for a given tibial width and can be used to predict the dimensions of a candidate knee based on a single measurement. The anatomical/anthropometric data presented in the study can be utilized in a sizing algorithm for artificial meniscal implants or in the design of artificial meniscus prostheses.
    • Download: (899.1Kb)
    • Show Full MetaData Hide Full MetaData
    • Get RIS
    • Item Order
    • Go To Publisher
    • Statistics

      MRI-Based Characterization of Bone Anatomy in the Human Knee for Size Matching of a Medial Meniscal Implant

    URI
    https://yetl.yabesh.ir/yetl1/handle/yetl/142536
    Collections
    • Journal of Biomechanical Engineering

    Show full item record

    contributor authorJonathan J. Elsner
    contributor authorFarshid Guilak
    contributor authorAvi Shterling
    contributor authorEran Linder-Ganz
    contributor authorSigal Portnoy
    date accessioned2017-05-09T00:36:27Z
    date available2017-05-09T00:36:27Z
    date copyrightOctober, 2010
    date issued2010
    identifier issn0148-0731
    identifier otherJBENDY-27171#101008_1.pdf
    identifier urihttp://yetl.yabesh.ir/yetl/handle/yetl/142536
    description abstractAllograft or synthetic menisci have been suggested as a means to restore contact pressures following meniscectomy. However, when the natural meniscus is severely damaged/absent, the necessary size cannot be determined according to the recipient size and there is a need to estimate it from magnetic resonance imaging (MRI) of the contralateral knee or the injured knee bones. The use of the contralateral-knee for size matching is problematic due to economic and practical reasons. Hence, there are significant advantages for a sizing algorithm based only on the candidate knee geometry. The aim of this study is to characterize midrange values and variability of knee dimensions and to develop a set of mathematical relations representing knee dimensions using a minimum of imaging-based bone measurements. Tibia, femur, and meniscus measurements were taken in 118 MRI scans and used to develop a representative parametric knee model in which all dimensions are expressed using tibia plateau width. The model was verified by comparing the predicted values to direct MRI measurements for 20 additional subjects by means of the Pearson correlation and and (1986, “Statistical Methods for Assessing Agreement Between Two Methods of Clinical Measurement,” Lancet, 1, pp. 307–310) plot. Anatomical parameters in the male knee were significantly larger (∼17%) compared with corresponding female measurements. However, most relations between tibia, femur, and meniscus measurements (43/56) were not significantly different between male and female populations (p≥0.05), indicating that differences between male and female joints are generally related to scaling and not shape. Dimensions predicted by the knee model were in a good agreement with dimensions measured directly from the MRI (R2>0.96) and the Bland and Altman plot indicated that ∼95% of data points were well within the ±2 standard deviation lines of agreement. The model proposed in this study is advantageous in being able to describe typical knee proportions for a given tibial width and can be used to predict the dimensions of a candidate knee based on a single measurement. The anatomical/anthropometric data presented in the study can be utilized in a sizing algorithm for artificial meniscal implants or in the design of artificial meniscus prostheses.
    publisherThe American Society of Mechanical Engineers (ASME)
    titleMRI-Based Characterization of Bone Anatomy in the Human Knee for Size Matching of a Medial Meniscal Implant
    typeJournal Paper
    journal volume132
    journal issue10
    journal titleJournal of Biomechanical Engineering
    identifier doi10.1115/1.4002490
    journal fristpage101008
    identifier eissn1528-8951
    treeJournal of Biomechanical Engineering:;2010:;volume( 132 ):;issue: 010
    contenttypeFulltext
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian
     
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian