    A New System for Monitoring Gait Training in Infants With Down Syndrome

    Source: Journal of Medical Devices:;2009:;volume( 003 ):;issue: 002::page 24501
    Author:
    Sigal Portnoy
    ,
    Ziva Yizhar
    ,
    Amit Gefen
    ,
    Assaf Ohana
    DOI: 10.1115/1.3116251
    Publisher: The American Society of Mechanical Engineers (ASME)
    Abstract: Infants with Down syndrome show an altered pattern of motor development at early childhood. Treadmill-walking training can be used to promote the earlier attainment of motor milestones in infants with locomotion deficiencies but quantitative data on their motor development are limited to gait laboratory studies. Our purpose was to develop, validate, and test a low-cost portable system for detecting infant steps on a treadmill while training. The system includes five infrared distance sensors, which were placed on a motorized treadmill to record the location of the feet of the infant during walking and thus measure his/her step length and cadence. The system was validated using synthetic objects, with a healthy 13-month-old infant. Pilot studies were then conducted in additional five infants with Down syndrome (aged 11–28 months) to determine step length (17.5–22.3 cm) and cadence (0.33–2.16 steps/s) at baseline, as well as at follow-ups 1 month and 3 months after the first trial. Measurements were repeatable per session and agreed with values reported in the literature. These pilot studies indicate the potential utility of the present system in quantitative monitoring of the process of acquisition of initial gait in infants with Down syndrome at the care facility where routine therapy is given.
    • Download: (276.8Kb)
    • Show Full MetaData Hide Full MetaData
    • Get RIS
    • Item Order
    • Go To Publisher
    • Statistics

      A New System for Monitoring Gait Training in Infants With Down Syndrome

    URI
    https://yetl.yabesh.ir/yetl1/handle/yetl/141648
    Collections
    • Journal of Medical Devices

    Show full item record

    contributor authorSigal Portnoy
    contributor authorZiva Yizhar
    contributor authorAmit Gefen
    contributor authorAssaf Ohana
    date accessioned2017-05-09T00:34:47Z
    date available2017-05-09T00:34:47Z
    date copyrightJune, 2009
    date issued2009
    identifier issn1932-6181
    identifier otherJMDOA4-28002#024501_1.pdf
    identifier urihttp://yetl.yabesh.ir/yetl/handle/yetl/141648
    description abstractInfants with Down syndrome show an altered pattern of motor development at early childhood. Treadmill-walking training can be used to promote the earlier attainment of motor milestones in infants with locomotion deficiencies but quantitative data on their motor development are limited to gait laboratory studies. Our purpose was to develop, validate, and test a low-cost portable system for detecting infant steps on a treadmill while training. The system includes five infrared distance sensors, which were placed on a motorized treadmill to record the location of the feet of the infant during walking and thus measure his/her step length and cadence. The system was validated using synthetic objects, with a healthy 13-month-old infant. Pilot studies were then conducted in additional five infants with Down syndrome (aged 11–28 months) to determine step length (17.5–22.3 cm) and cadence (0.33–2.16 steps/s) at baseline, as well as at follow-ups 1 month and 3 months after the first trial. Measurements were repeatable per session and agreed with values reported in the literature. These pilot studies indicate the potential utility of the present system in quantitative monitoring of the process of acquisition of initial gait in infants with Down syndrome at the care facility where routine therapy is given.
    publisherThe American Society of Mechanical Engineers (ASME)
    titleA New System for Monitoring Gait Training in Infants With Down Syndrome
    typeJournal Paper
    journal volume3
    journal issue2
    journal titleJournal of Medical Devices
    identifier doi10.1115/1.3116251
    journal fristpage24501
    identifier eissn1932-619X
    treeJournal of Medical Devices:;2009:;volume( 003 ):;issue: 002
    contenttypeFulltext
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian
     
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian