    Impacts of Wireless Power on Medical Device Design Safety

    Source: Journal of Medical Devices:;2009:;volume( 003 ):;issue: 002::page 27544
    Author:
    J. Schneider
    ,
    L. Lucke
    ,
    D. Wessels
    ,
    T. Schauer
    DOI: 10.1115/1.3135198
    Publisher: The American Society of Mechanical Engineers (ASME)
    Abstract: Wireless power holds great promise for solving many power distribution problems. Medical device designers will need to understand the impact of the electromagnetic coupling used for wireless power systems to design safe electromagnetic environments and safe medical devices. One question for designers will be whether or not current standards and requirements used for testing the electromagnetic compatibility (EMC) of medical devices and human exposure go far enough to insure safe environments and safe and reliable medical devices in the presence of wireless power. Electromagnetic energy can be transferred in three ways: through induction, radio frequency waves, or resonant evanescent coupling. Nonradiative inductive coupling uses the magnetic fields created when current is passed through one coil to create a current in a second coil that is located very near the first coil. These systems usually operate in the 50 KHz to 10 MHz range. Radio frequency energy can be transferred through radiating electromagnetic waves over great distances at frequencies from the upper KHz to many GHz. Most recently, work has been done on resonant evanescent coupling which transfers power between resonant objects over a distance of a couple of meters at frequencies from 1–10 MHz. Safety and reliability of medical devices is confirmed by testing EMC emissions and susceptibility to IEC60601-1-2 and supporting standards. For example, one of the supporting standards, CISPR 11 calls for measuring the electric field of radiated emissions over 30 MHz and the magnetic field below 30 MHz at distances of 3–10 meters. Many of the effects of wireless power systems are in the near field and are not covered in the current test standards. The AAMI PC69 series of standards have some near field requirements but these standards tend to be industry specific – such as drug pumps or pacemakers. EMC immunity standards used to test EMC susceptibility barely mention magnetic immunity. The only test for magnetic fields recommends testing fields at the power frequencies of 50 and 60 Hz. There are few standards detailing safe limits for human exposure to the near field effects of wireless power as well. Historically human exposure standards have been based on time average thermal effects on tissue and not medical devices. IEEE's C95.1b has requirements for specific absorption rate limits averaged over a 6 minute period. A pulsed wireless power system could meet these requirements and be safe for exposed tissue, but if a patient has an implanted device, or is wearing an external medical device, the pulsed EM energy could affect it during the pulse. The German BGV B11 standard lists human exposure limits for electric and magnetic fields based on a time average and limits exposure based on which portion of the body is exposed. However, it is meant as a workplace standard not a medical device standard. Currently the FDA does not require meeting either of these standards. It is necessary to determine the appropriate limits and tests to ensure that medical devices safely use wireless power and continue to operate safely in the presence of wireless power.
    • Download: (37.87Kb)
    • Show Full MetaData Hide Full MetaData
    • Get RIS
    • Item Order
    • Go To Publisher
    • Statistics

      Impacts of Wireless Power on Medical Device Design Safety

    URI
    https://yetl.yabesh.ir/yetl1/handle/yetl/141633
    Collections
    • Journal of Medical Devices

    Show full item record

    contributor authorJ. Schneider
    contributor authorL. Lucke
    contributor authorD. Wessels
    contributor authorT. Schauer
    date accessioned2017-05-09T00:34:46Z
    date available2017-05-09T00:34:46Z
    date copyrightJune, 2009
    date issued2009
    identifier issn1932-6181
    identifier otherJMDOA4-28002#027544_1.pdf
    identifier urihttp://yetl.yabesh.ir/yetl/handle/yetl/141633
    description abstractWireless power holds great promise for solving many power distribution problems. Medical device designers will need to understand the impact of the electromagnetic coupling used for wireless power systems to design safe electromagnetic environments and safe medical devices. One question for designers will be whether or not current standards and requirements used for testing the electromagnetic compatibility (EMC) of medical devices and human exposure go far enough to insure safe environments and safe and reliable medical devices in the presence of wireless power. Electromagnetic energy can be transferred in three ways: through induction, radio frequency waves, or resonant evanescent coupling. Nonradiative inductive coupling uses the magnetic fields created when current is passed through one coil to create a current in a second coil that is located very near the first coil. These systems usually operate in the 50 KHz to 10 MHz range. Radio frequency energy can be transferred through radiating electromagnetic waves over great distances at frequencies from the upper KHz to many GHz. Most recently, work has been done on resonant evanescent coupling which transfers power between resonant objects over a distance of a couple of meters at frequencies from 1–10 MHz. Safety and reliability of medical devices is confirmed by testing EMC emissions and susceptibility to IEC60601-1-2 and supporting standards. For example, one of the supporting standards, CISPR 11 calls for measuring the electric field of radiated emissions over 30 MHz and the magnetic field below 30 MHz at distances of 3–10 meters. Many of the effects of wireless power systems are in the near field and are not covered in the current test standards. The AAMI PC69 series of standards have some near field requirements but these standards tend to be industry specific – such as drug pumps or pacemakers. EMC immunity standards used to test EMC susceptibility barely mention magnetic immunity. The only test for magnetic fields recommends testing fields at the power frequencies of 50 and 60 Hz. There are few standards detailing safe limits for human exposure to the near field effects of wireless power as well. Historically human exposure standards have been based on time average thermal effects on tissue and not medical devices. IEEE's C95.1b has requirements for specific absorption rate limits averaged over a 6 minute period. A pulsed wireless power system could meet these requirements and be safe for exposed tissue, but if a patient has an implanted device, or is wearing an external medical device, the pulsed EM energy could affect it during the pulse. The German BGV B11 standard lists human exposure limits for electric and magnetic fields based on a time average and limits exposure based on which portion of the body is exposed. However, it is meant as a workplace standard not a medical device standard. Currently the FDA does not require meeting either of these standards. It is necessary to determine the appropriate limits and tests to ensure that medical devices safely use wireless power and continue to operate safely in the presence of wireless power.
    publisherThe American Society of Mechanical Engineers (ASME)
    titleImpacts of Wireless Power on Medical Device Design Safety
    typeJournal Paper
    journal volume3
    journal issue2
    journal titleJournal of Medical Devices
    identifier doi10.1115/1.3135198
    journal fristpage27544
    identifier eissn1932-619X
    treeJournal of Medical Devices:;2009:;volume( 003 ):;issue: 002
    contenttypeFulltext
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian
     
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian