    Design Innovation Size and Shape Optimization of a 1.0mm Multifunctional Forceps-Scissors Surgical Instrument

    Source: Journal of Medical Devices:;2008:;volume( 002 ):;issue: 001::page 15001
    Author:
    Milton E. Aguirre
    ,
    Mary Frecker
    DOI: 10.1115/1.2885141
    Publisher: The American Society of Mechanical Engineers (ASME)
    Abstract: A size and shape optimization routine is developed for a 1.0mm diameter multifunctional instrument for minimally invasive surgery. The instrument is a compliant mechanism capable of both grasping and cutting. Multifunctional instruments are expected to be beneficial in the operating room because of their ability to perform multiple surgical tasks, thereby decreasing the total number of instrument exchanges in a single procedure. With fewer instrument exchanges, the risk of inadvertent tissue trauma as well as overall surgical time and costs are reduced. The focus of this paper is to investigate the performance effects of allowing the cross-sectional area along the length of the device to vary. This investigation is accomplished by defining various cross-sectional segments in terms of parametric variables and optimizing the dimensions of the instrument to provide a sufficient opening of the forceps jaws while maintaining adequate cutting and grasping forces. Two optimization problems are considered. First, all parametric segments are set equal to one another to achieve size optimization. Second, each segment is allowed to vary independently, thereby achieving shape optimization. Large deformation finite element analysis and optimization are conducted using ANSYS®. Finally, prototypes are fabricated using wire EMD and experiments are conducted to evaluate the instrument performance. As a result of allowing the cross-sectional area to vary, i.e., conducting shape optimization, the forceps and scissors blocked forces increased by as much as 83.2% and 87%, respectively. During prototype evaluations, it is found that the finite element analysis predictions were within 10% of the measured tool performance. Therefore, for this application, it is concluded that performing shape optimization does significantly influence the performance of the instrument.
    keyword(s): Force , Engineering prototypes , Design , Instrumentation , Optimization , Shapes , Grasping , Cutting AND Dimensions ,
    • Download: (852.9Kb)
    • Show Full MetaData Hide Full MetaData
    • Get RIS
    • Item Order
    • Go To Publisher
    • Statistics

      Design Innovation Size and Shape Optimization of a 1.0mm Multifunctional Forceps-Scissors Surgical Instrument

    URI
    https://yetl.yabesh.ir/yetl1/handle/yetl/139085
    Collections
    • Journal of Medical Devices

    Show full item record

    contributor authorMilton E. Aguirre
    contributor authorMary Frecker
    date accessioned2017-05-09T00:30:02Z
    date available2017-05-09T00:30:02Z
    date copyrightMarch, 2008
    date issued2008
    identifier issn1932-6181
    identifier otherJMDOA4-27988#015001_1.pdf
    identifier urihttp://yetl.yabesh.ir/yetl/handle/yetl/139085
    description abstractA size and shape optimization routine is developed for a 1.0mm diameter multifunctional instrument for minimally invasive surgery. The instrument is a compliant mechanism capable of both grasping and cutting. Multifunctional instruments are expected to be beneficial in the operating room because of their ability to perform multiple surgical tasks, thereby decreasing the total number of instrument exchanges in a single procedure. With fewer instrument exchanges, the risk of inadvertent tissue trauma as well as overall surgical time and costs are reduced. The focus of this paper is to investigate the performance effects of allowing the cross-sectional area along the length of the device to vary. This investigation is accomplished by defining various cross-sectional segments in terms of parametric variables and optimizing the dimensions of the instrument to provide a sufficient opening of the forceps jaws while maintaining adequate cutting and grasping forces. Two optimization problems are considered. First, all parametric segments are set equal to one another to achieve size optimization. Second, each segment is allowed to vary independently, thereby achieving shape optimization. Large deformation finite element analysis and optimization are conducted using ANSYS®. Finally, prototypes are fabricated using wire EMD and experiments are conducted to evaluate the instrument performance. As a result of allowing the cross-sectional area to vary, i.e., conducting shape optimization, the forceps and scissors blocked forces increased by as much as 83.2% and 87%, respectively. During prototype evaluations, it is found that the finite element analysis predictions were within 10% of the measured tool performance. Therefore, for this application, it is concluded that performing shape optimization does significantly influence the performance of the instrument.
    publisherThe American Society of Mechanical Engineers (ASME)
    titleDesign Innovation Size and Shape Optimization of a 1.0mm Multifunctional Forceps-Scissors Surgical Instrument
    typeJournal Paper
    journal volume2
    journal issue1
    journal titleJournal of Medical Devices
    identifier doi10.1115/1.2885141
    journal fristpage15001
    identifier eissn1932-619X
    keywordsForce
    keywordsEngineering prototypes
    keywordsDesign
    keywordsInstrumentation
    keywordsOptimization
    keywordsShapes
    keywordsGrasping
    keywordsCutting AND Dimensions
    treeJournal of Medical Devices:;2008:;volume( 002 ):;issue: 001
    contenttypeFulltext
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian
     
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian