    Flexure-Based Locking Plates Can Modulate Interfragmentary Motion in Distal Femur and Diaphyseal Fractures: A Parametric Finite Element Analysis

    Thumbnail
    View/Open
    bio-25-1157.pdf (2.750Mb)
    Date
    2026
    Author
    Huxman, Connor
    Updegrove, Gary
    Armstrong, April
    Wee, Hwabok
    Frecker, Mary I.
    Butler, Jared
    Lewis, Gregory S.
    Metadata
    Show full item record
    Abstract
    Abstract. Axial interfragmentary motion is known to stimulate fracture healing. A mechanically compliant fracture fixation plate incorporating flexures is proposed to provide controlled axial micromotion to long bone fractures. To explore the concept's feasibility, computational modeling of general diaphyseal and distal femur fractures treated with both rigid and compliant plates is conducted. In Part I of this study, a diaphyseal fracture finite element model for novel compliant plates is validated against experimental data with good agreement. In Part II, a parametric analysis is conducted using the validated model to characterize the performance of many compliant plate designs with varying geometry and materials. Under axial loading, all compliant plate configurations provided greater (1.03 mm versus 0.22 mm) and more symmetric (270–390%) axial interfragmentary motion than rigid plates. Steel compliant plates with thicker flexures (0.3–0.6 mm) may provide the best performance given their enhanced motion and comparable bending/torsional rigidity. In Part III, compliant plates are adapted for use in treating distal femur fractures. Results demonstrate that compared to a rigid plate, a compliant distal femur plate with increased thickness can effectively modulate interfragmentary motion—that is, increase the insufficient near cortex motion under low loads (from 0.14 mm to 0.23 mm) and reduce the excessive far cortex motion under large loads (from 7.96 mm to 2.54 mm). Flexure-based locking plates represent a promising new approach to treating diaphyseal and/or distal femur fractures. Additional research is needed to investigate in vivo performance.
    URI
    http://yetl.yabesh.ir/yetl1/handle/yetl/4316673
    Collections
    • Journal of Biomechanical Engineering

    DSpace software copyright © 2002-2015  DuraSpace
    Contact Us | Send Feedback
    Theme by 
    @mire NV
     

     

    Browse

    All of DSpaceCommunities & CollectionsBy Issue DateAuthorsTitlesSubjectsThis CollectionBy Issue DateAuthorsTitlesSubjects

    My Account

    LoginRegister

    Statistics

    View Usage Statistics

    DSpace software copyright © 2002-2015  DuraSpace
    Contact Us | Send Feedback
    Theme by 
    @mire NV