    Temperature and Viscosity Effects on Self-Expanding Nitinol Transcatheter Aortic Valve Performance

    Thumbnail
    View/Open
    med-25-1276.pdf (1.073Mb)
    Date
    2026
    Author
    Bshennaty, Ahmad
    Vogl, Brennan
    Zhang, Zhongtian
    Saleh, Ghasaq
    Lee, Bruce
    Alkhouli, Mohamad
    Hatoum, Hoda
    Metadata
    Show full item record
    Abstract
    Abstract. In Vitro testing is widely used to evaluate the hemodynamic performance of transcatheter aortic valves (TAVs). However, many studies assess self-expanding valves using blood analogs matched in viscosity at room temperature, despite the temperature-dependent mechanical behavior of nitinol stent frames. This study investigates the independent and combined effects of temperature and viscosity on the in vitro performance assessment of a self-expanding TAV. Rheological characterization was performed to identify water-glycerin solutions with matched viscosities at room temperature and physiological temperature (37 °C). A patient-specific aortic model was cast in silicone. A self-expanding Medtronic Evolut R valve was deployed within the compliant model and evaluated using a pulse-duplicating left heart simulator under physiological flow and pressure conditions. The transvalvular pressure gradient and effective orifice area (EOA) were calculated. Rheometry identified two fluid conditions with comparable viscosities at room and body temperatures. Heating the blood analog to 37 °C resulted in significantly reduced transvalvular pressure gradient and increased EOA compared to the viscosity-matched room temperature condition (p < 0.0001). At physiological body temperature, the higher viscosity fluid (42% glycerin) yielded further improvements in valve performance relative to the lower viscosity fluid (38% glycerin), with lower pressure drop and larger EOA (p < 0.0001). These results demonstrate that temperature-dependent effects influence the hemodynamic performance of self-expanding nitinol TAVs beyond viscosity matching alone. Combined consideration of temperature and viscosity is therefore necessary to ensure physiologically relevant in vitro assessment of self-expanding TAVs.
    URI
    http://yetl.yabesh.ir/yetl1/handle/yetl/4315589
    Collections
    • Journal of Medical Devices

    DSpace software copyright © 2002-2015  DuraSpace
    Contact Us | Send Feedback
    Theme by 
    @mire NV
     

     

    Browse

    All of DSpaceCommunities & CollectionsBy Issue DateAuthorsTitlesSubjectsThis CollectionBy Issue DateAuthorsTitlesSubjects

    My Account

    LoginRegister

    Statistics

    View Usage Statistics

    DSpace software copyright © 2002-2015  DuraSpace
    Contact Us | Send Feedback
    Theme by 
    @mire NV