• Login
    View Item 
    •   DSpace Home
    • ASME
    • Journal of Medical Devices
    • View Item
    •   DSpace Home
    • ASME
    • Journal of Medical Devices
    • View Item
    JavaScript is disabled for your browser. Some features of this site may not work without it.

    Intraoperative Dual-Cantilever Probe for Rapid Measurement of Cancellous Bone Compressive Strength Within the Vertebral Body

    Thumbnail
    View/Open
    med-25-1069.pdf (1.348Mb)
    Date
    2026
    Author
    Kresse, Owen
    Khong, Evelyn
    Loohuis, Gerhardus O.
    Elayne Shelby, R.
    Boakye, Maxwell
    Voor, Michael J.
    Williams, Stuart J.
    Metadata
    Show full item record
    Abstract
    Abstract. Dual X-ray absorptiometry (DEXA) scans are the current standard in assessing bone mineral density (BMD) and are used to identify patients who may need screw augmentation during spinal fusion. DEXA scans are not always available and tend to overestimate BMD. This paper describes the development of a dual-cantilever mechanical probe and tested in polyurethane foam blocks with varying compressive strengths. The probe was modeled after a 5.5 mm tap and foam block holes were prepared replicating intra-operative conditions. Calibration curves were acquired for each probe using six foam blocks (1.5–12.9 MPa). Verification tests were performed in a different set of four foam blocks (2.05–9.65 MPa). Four probes were machined and tested for repeatability. Three users separately acquired measurements of foam blocks to test for reliability. The root-mean-square error of all four probes measuring the 2.05 MPa, 3.65 MPa, 5.80 MPa, and 9.65 MPa samples were 0.89 MPa, 0.32 MPa, 1.41 MPa, and 1.71 MPa, respectively. There was not a significant difference between different probes or different users. The dual-cantilever probe provided measurements within the clinically relevant range of compressive strengths for vertebral trabecular bone. A targeted and reliable bone strength measurement technique could reduce the occurrence and revision surgeries and improve patient outcomes.
    URI
    http://yetl.yabesh.ir/yetl1/handle/yetl/4315573
    Collections
    • Journal of Medical Devices

    DSpace software copyright © 2002-2015  DuraSpace
    Contact Us | Send Feedback
    Theme by 
    @mire NV
     

     

    Browse

    All of DSpaceCommunities & CollectionsBy Issue DateAuthorsTitlesSubjectsThis CollectionBy Issue DateAuthorsTitlesSubjects

    My Account

    LoginRegister

    Statistics

    View Usage Statistics

    DSpace software copyright © 2002-2015  DuraSpace
    Contact Us | Send Feedback
    Theme by 
    @mire NV