    The Effect of Over-Rotating Nickel–Titanium Medical Retrieval Device Guidewires—Part I: Key Parameters Causing Kink Formation Resulting in Premature and Catastrophic Failure

    Thumbnail
    View/Open
    med-25-1113.pdf (1.697Mb)
    Date
    2026
    Author
    Puttlitz, Karl J.
    Frakes, David
    Metadata
    Show full item record
    Abstract
    Abstract. A variety of Nitinol-based endovascular mechanical devices have been utilized to capture and retrieve blood clots and foreign objects such as embolic coils that occasionally escape during aneurysm treatments and other interventions throughout the vascular system. Nitinol is one of a few materials possessing a novel characteristic referred to as shape memory alloy (SMA). This characteristic allows the material to be deformed, but immediately spring back to its original shape with moderate heating, hence the term “shape memory.” Nitinol, however, also exhibits a well-documented concern when used in guidewires, i.e., a tendency to form cracks transverse to the longitudinal wire axis on the compression side of a kink that results from a loop created by over rotating the guidewire. The cracks can then propagate through the wire causing a dramatic loss of strength accompanied by premature and catastrophic failure with only a minor applied force. It was a key objective of the research described herein to determine the critical catheter-to-attached appliance distance (CAAD) at which and below a loop would not form during a procedure. It was a further objective to determine the measured loop diameter (MLD) of a kink that results in transverse cracks on the compression side that causes premature and catastrophic guidewire failure and contributing factors for loop formation. This paper also defines design guidelines for ways to prevent the ill effects of loop formation by immediately becoming aware of their existence and then eliminating them before reaching the critical diameter.
    URI
    http://yetl.yabesh.ir/yetl1/handle/yetl/4315555
    Collections
    • Journal of Medical Devices

    DSpace software copyright © 2002-2015  DuraSpace
    Contact Us | Send Feedback
    Theme by 
    @mire NV
     

     

    Browse

    All of DSpaceCommunities & CollectionsBy Issue DateAuthorsTitlesSubjectsThis CollectionBy Issue DateAuthorsTitlesSubjects

    My Account

    LoginRegister

    Statistics

    View Usage Statistics

    DSpace software copyright © 2002-2015  DuraSpace
    Contact Us | Send Feedback
    Theme by 
    @mire NV