    Performance of a Dual-Sensor Cardiopulmonary Resuscitation Feedback System in Estimating Compression Depth and Rate During Simulated Chest Compressions

    Thumbnail
    View/Open
    med_019_02_021009.pdf (1.524Mb)
    Date
    2025
    Author
    Gould, Jeffrey R.
    Mostofi, Naghmeh
    Badin, Jill K.
    Giacometti, Paolo
    Lampe, Joshua W.
    Metadata
    Show full item record
    Abstract
    This study evaluated the performance of a dual-sensor cardiopulmonary resuscitation (CPR) feedback system in estimating chest compression depth and rate over a range of depth and rate combinations during rigid and compliant surface conditions using a computer-controlled motion system to simulate chest compressions. Ten dual-sensor CPR pads were tested using a computer-controlled motion system which simulated chest compressions at target depths of 1.9, 3.8, 4.8, 6.4, and 8.9 cm and target rates of 60, 80, 100, 120, and 140 compressions per minute (cpm). A rigid surface was simulated by applying motion only to the anterior sensor, and a compliant surface was simulated by applying motion to both the anterior and posterior sensor, challenging the algorithm to calculate a net compression depth by subtracting motion of the posterior sensor. For all simulated compressions, including every rate and depth combination and for both rigid and compliant surface simulations, the mean (±sd) depth error was 0.05 (±0.08) cm and the rate error was −0.55 (±1.44) cpm. A dual-sensor CPR system accurately estimates compression depth within ±0.25 cm and compression rate within ±3 cpm over a wide range and combination of clinically relevant chest compression depths and rates during both rigid and compliant surface simulations. Use of a computer-controlled motion system provides a more direct assessment of accuracy than manual compressions performed on instrumented manikins.
    URI
    http://yetl.yabesh.ir/yetl1/handle/yetl/4308079
    Collections
    • Journal of Medical Devices

    DSpace software copyright © 2002-2015  DuraSpace
    Contact Us | Send Feedback
    Theme by 
    @mire NV
     

     

    Browse

    All of DSpaceCommunities & CollectionsBy Issue DateAuthorsTitlesSubjectsThis CollectionBy Issue DateAuthorsTitlesSubjects

    My Account

    LoginRegister

    Statistics

    View Usage Statistics

    DSpace software copyright © 2002-2015  DuraSpace
    Contact Us | Send Feedback
    Theme by 
    @mire NV