Show simple item record

dc.contributor.authorCeco, Irma
dc.contributor.authorYanez, Leonardo Zamora
dc.contributor.authorScott, Peter
dc.contributor.authorLi, Mo
dc.contributor.authorSlocum, Alexander
dc.contributor.authorFuenfer, Michael
dc.date.accessioned2024-12-24T19:14:52Z
dc.date.available2024-12-24T19:14:52Z
dc.date.copyright7/30/2024 12:00:00 AM
dc.date.issued2024
dc.identifier.issn1932-6181
dc.identifier.othermed_018_03_035001.pdf
dc.identifier.urihttp://yetl.yabesh.ir/yetl1/handle/yetl/4303577
dc.description.abstractThe majority of hospitalized patients require the insertion of a catheter as part of their medical treatment. For catheters to be connected to external devices, barbed connectors are often employed. If this connection is ever disrupted, a direct channel from the environment to the patient's internal body may be introduced, leading to a life-threatening emergency. Current methods of securing catheter tubes to connectors have various limitations that have prevented their widespread adoption in medical settings, ranging from low ease of use to poor geometric compatibility. The authors present a device, called the Catheter Connector Clamp (C-3), shown in Fig. 1 that will securely clamp multisize catheter tubes to existing barbed connectors to increase the security and repeatability of the connection. The authors also propose a method for comparing the effectiveness of current catheter connection solutions to their C-3 design using maximum pull-off force. On silicone medical tubing, the C-3 clamp performed more than twice as well as friction alone (2.09 times more), and 25% better than the hospital gold-standard taped connection.
dc.publisherThe American Society of Mechanical Engineers (ASME)
dc.titleA C-3 Clamp for Improved Connection Between Catheter Tube and Barbed Connector Together With Its Evaluation Using a Maximum Pull-Off Force Method
dc.typeJournal Paper
dc.journal.volume18
dc.journal.issue3
dc.journal.titleJournal of Medical Devices
dc.identifier.doi10.1115/1.4065811
dc.journal.fristpage35001-1
dc.journal.lastpage35001-8
dc.page8
dc.treeJournal of Medical Devices:;2024:;volume( 018 ):;issue: 003
dc.contenttypeFulltext


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record