Show simple item record

dc.contributor.authorKnutson, Robert G.
dc.contributor.authorWhitten, Justin
dc.contributor.authorGraham, David
dc.contributor.authorShankwitz, Craig
dc.contributor.authorPew, Corey A.
dc.date.accessioned2024-12-24T19:14:28Z
dc.date.available2024-12-24T19:14:28Z
dc.date.copyright5/30/2024 12:00:00 AM
dc.date.issued2024
dc.identifier.issn1932-6181
dc.identifier.othermed_018_02_021005.pdf
dc.identifier.urihttp://yetl.yabesh.ir/yetl1/handle/yetl/4303565
dc.description.abstractTrips and falls are a major concern for older adults. The resulting injury and loss of mobility can have a significant impact on quality of life. An emerging field of study, known as Perturbation Training, has been shown to reduce injury rates associated with trips and falls in older adults. Treadmills traditionally used for Perturbation Training are large, expensive, and immobile, forcing users to travel long distances to receive care. A portable treadmill would serve a larger portion of the at-risk population than current methods. We developed a portable, low-cost, twin-belt perturbation treadmill capable of high-intensity Perturbation Training. Belt speeds are controlled by a custom mechanical and software interface, allowing operators with no programming experience to control the device. The treadmill can accommodate users up to 118 kg and provides a maximum acceleration and speed of 12 m/s2 and 3.3 m/s, respectively, under full load. The total weight is 180 kg, and the treadmill can be moved like a wheelbarrow, with handles in the back and wheels in the front. The prototype was validated with mechanical and human participant testing, showing it as a viable device for Perturbation Training. In this paper, we will go over the design, fabrication, and validation processes used to create the Portable Perturbation Treadmill.
dc.publisherThe American Society of Mechanical Engineers (ASME)
dc.titleDesign, Fabrication, and Validation of a Portable Perturbation Treadmill for Balance Recovery Research
dc.typeJournal Paper
dc.journal.volume18
dc.journal.issue2
dc.journal.titleJournal of Medical Devices
dc.identifier.doi10.1115/1.4065514
dc.journal.fristpage21005-1
dc.journal.lastpage21005-6
dc.page6
dc.treeJournal of Medical Devices:;2024:;volume( 018 ):;issue: 002
dc.contenttypeFulltext


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record