Show simple item record

dc.contributor.authorLi, Jihua
dc.contributor.authorJiang, Qi
dc.date.accessioned2023-08-16T18:45:05Z
dc.date.available2023-08-16T18:45:05Z
dc.date.copyright4/17/2023 12:00:00 AM
dc.date.issued2023
dc.identifier.issn1932-6181
dc.identifier.othermed_017_02_021008.pdf
dc.identifier.urihttp://yetl.yabesh.ir/yetl1/handle/yetl/4292434
dc.description.abstractIn this paper, a cable-driven puncturing surgery robot named cable-driven puncturing surgery robot (CPSR) is proposed for soft needle. The cable-driven mechanism has lower mass, smaller dimension and smooth transmission. The motor driving the rotational joints is separated using cable, which strengthens the dynamic performance of the robot. A decoupling mechanism based on movable pulley is also proposed to eliminate the cable coupling error. Transmission error of cable-driven mechanism is also analyzed to improve the control accuracy. The simulation of transmission error is completed to optimize the design parameters of cable-driven mechanism. In experiments, the feasibility of the cable decoupling mechanism is validated. Also the insertion error less than 0.7 mm and repeatability less than 2.5 mm are achieved. Finally the reasons limiting the accuracy and corresponding solutions are discussed. In the future research, the shape sensor will be fabricated using optic fiber and mounted on the slim needle which are helpful to complete automated operations with higher accuracy.
dc.publisherThe American Society of Mechanical Engineers (ASME)
dc.titleOptimal Design and Experiment of Cable-Driven Puncturing Surgery Robot for Soft Needle
dc.typeJournal Paper
dc.journal.volume17
dc.journal.issue2
dc.journal.titleJournal of Medical Devices
dc.identifier.doi10.1115/1.4056865
dc.journal.fristpage21008-1
dc.journal.lastpage21008-12
dc.page12
dc.treeJournal of Medical Devices:;2023:;volume( 017 ):;issue: 002
dc.contenttypeFulltext


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record