    An Experimental Study of Feasibility of a Mini-Invasive Fixator for Rib Osteosynthesis

    Thumbnail
    View/Open
    med_017_01_011001.pdf (1.508Mb)
    Date
    2022
    Author
    Puglisi, Lucrezia
    Ceccarelli, Marco
    Ambrogi, Vicenzo
    Metadata
    Show full item record
    Abstract
    Alternatives to the current invasive rib implants are discussed as from the current solutions with their constraints and disadvantages mainly in surgery application and clinical consequences. A novel rib fixator is presented with an experimental validation and characterization for a potential implementation in facilitating osteosynthesis of multifractured ribs. Testing is designed with lab facilities replicating normal breathing with the possibility of simulating cough too. Results of lab test with a ribolution rib fixator prototype are discussed from experiences with pig ribs with satisfactory results in behavior and numerical values.