Show simple item record

dc.contributor.authorMehta
dc.contributor.authorJason N.;Morales
dc.contributor.authorBrianna E.;Hsu
dc.contributor.authorFang-Chi;Rossmeisl
dc.contributor.authorJohn H.;Rylander
dc.contributor.authorChristopher G.
dc.date.accessioned2022-08-18T12:54:31Z
dc.date.available2022-08-18T12:54:31Z
dc.date.copyright6/16/2022 12:00:00 AM
dc.date.issued2022
dc.identifier.issn0148-0731
dc.identifier.otherbio_144_11_111003.pdf
dc.identifier.urihttp://yetl.yabesh.ir/yetl1/handle/yetl/4287078
dc.description.abstractConvection-enhanced delivery (CED) has been extensively studied for drug delivery to the brain due to its inherent ability to bypass the blood-brain barrier. Unfortunately, CED has also been shown to inadequately distribute therapeutic agents over a large enough targeted tissue volume to be clinically beneficial. In this study, we explore the use of constant pressure infusions in addition to controlled catheter movement as a means to increase volume dispersed (Vd) in an agarose gel brain tissue phantom. Constant flow rate and constant pressure infusions were conducted with a stationary catheter, a catheter retracting at a rate of 0.25 mm/min, and a catheter retracting at a rate of 0.5 mm/min. The 0.25 mm/min and 0.5 mm/min retracting constant pressure catheters resulted in significantly larger Vd compared to any other group, with a 105% increase and a 155% increase compared to the stationary constant flow rate catheter, respectively. These same constant pressure retracting infusions resulted in a 42% and 45% increase in Vd compared to their constant flow rate counterparts. Using constant pressure infusions coupled with controlled catheter movement appears to have a beneficial effect on Vd in agarose gel. Furthermore, constant pressure infusions reveal the fundamental limitation of flow-driven infusions in both controlled catheter movement protocols as well as in stationary protocols where maximum infusion volume can never be reliably obtained.
dc.publisherThe American Society of Mechanical Engineers (ASME)
dc.titleConstant Pressure Convection-Enhanced Delivery Increases Volume Dispersed With Catheter Movement in Agarose
dc.typeJournal Paper
dc.journal.volume144
dc.journal.issue11
dc.journal.titleJournal of Biomechanical Engineering
dc.identifier.doi10.1115/1.4054729
dc.journal.fristpage111003-1
dc.journal.lastpage111003-8
dc.page8
dc.treeJournal of Biomechanical Engineering:;2022:;volume( 144 ):;issue: 011
dc.contenttypeFulltext