    Scale Effects on Performance of BLDC Micromotors for Internal Biomedical Applications: A Finite Element Analysis

    Thumbnail
    View/Open
    med_016_03_031011.pdf (2.147Mb)
    Date
    2022
    Author
    Villalba-Alumbreros
    Gabriel;Moron-Alguacil
    Carlos;Fernandez-Munoz
    Miguel;Valiente-Blanco
    Ignacio;Diez-Jimenez
    Efren
    Metadata
    Show full item record
    Abstract
    Micromotors can be used to build up complex microtools for internal medical applications as, for example, steerable catheters or optical and ultrasonic imaging system. The thinner and smaller the micromotors are, the less invasive is the implantation. However, miniaturization of motors implies some limitations in torque, speed, and efficiency. This paper theoretically analyzes the scale effects on torque, efficiency, and thermal behavior of high torque permanent magnet brushless DC (BLDC) motors with ferromagnetic core coils operating in different in-body environment. Using a finite element model of a two-phase BLDC motor, scalability laws are provided for diameters between 0.1 and 100 mm and current densities between 1 and 1000 A/mm2. Based on the impact of the cogging torque and overheating of the motor, scale-dependent operational limits are calculated. Operational threshold can be determined at the point where cogging torque becomes dominating over total torque, limiting the use of traditional iron-core motors in the microscale. Current density limits are provided based on three representative in-body thermal scenarios: respiratory tract, body fluid, and blood torrent. Maximum current densities and corresponding torque and efficiency have been obtained for different micromotor sizes considering safe in-body temperature operation as threshold. It is demonstrated that micromotors of sizes down to 0.1 mm diameter could be used in internal body environments with acceptable performance.
    URI
    http://yetl.yabesh.ir/yetl1/handle/yetl/4286979
    Collections
    • Journal of Medical Devices

    DSpace software copyright © 2002-2015  DuraSpace
    Contact Us | Send Feedback
    Theme by 
    @mire NV
     

     

    Browse

    All of DSpaceCommunities & CollectionsBy Issue DateAuthorsTitlesSubjectsThis CollectionBy Issue DateAuthorsTitlesSubjects

    My Account

    LoginRegister

    Statistics

    View Usage Statistics

    DSpace software copyright © 2002-2015  DuraSpace
    Contact Us | Send Feedback
    Theme by 
    @mire NV