Show simple item record

dc.contributor.authorJoshi, Tanuj
dc.contributor.authorSharma, Ravikant
dc.contributor.authorMittal, Vinod Kumar
dc.contributor.authorGupta, Vikas
dc.contributor.authorKrishan, Gopal
dc.date.accessioned2022-05-08T08:58:10Z
dc.date.available2022-05-08T08:58:10Z
dc.date.copyright2/24/2022 12:00:00 AM
dc.date.issued2022
dc.identifier.issn2770-3495
dc.identifier.otheraoje_1_011001.pdf
dc.identifier.urihttp://yetl.yabesh.ir/yetl1/handle/yetl/4284569
dc.description.abstractIn the present paper, a three-dimensional finite element model of the Charnley implant has been developed to analyze the stress–strain distribution and deformation over the stem prosthesis. Patient-specific dynamic forces have been considered for the analytical evaluation using commercial finite element code. The impact of each dynamic activity has been analyzed separately using six different biocompatible alloy materials made of titanium and cobalt-chromium. Mechanical parameters have been evaluated to envisage the longevity and functionality of the implant. The performance of different materials for each suitable gait pattern is analyzed using finite element code. Consequently, Cobalt chromium alloys (CoCrMo alloy) demonstrate better results, i.e., maximum stress, minimum deformation, and strain as compared with other materials. Every dynamic motion, viz., walking, standing up, sitting down, going upstairs, and going downstairs are found in good agreement with the safety factor for every biomaterial. Additionally, going downstairs and sitting down gait motion exhibits the maximum and minimum stress–strain level, respectively. Based on the outcome of the presurgical study, it is recommended that CoCrMo alloys should be preferred over other materials.
dc.publisherThe American Society of Mechanical Engineers (ASME)
dc.titleDynamic Analysis of Hip Prosthesis Using Different Biocompatible Alloys
dc.typeJournal Paper
dc.journal.volume1
dc.journal.titleASME Open Journal of Engineering
dc.identifier.doi10.1115/1.4053417
dc.journal.fristpage11001-1
dc.journal.lastpage11001-10
dc.page10
dc.treeASME Open Journal of Engineering:;2022:;volume( 001 )
dc.contenttypeFulltext


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record