    Boundary Conditions Matter—Impact of Test Setup on Inferred Construct Mechanics in Plated Distal Femur Osteotomies

    Thumbnail
    View/Open
    bio_144_08_081009.pdf (2.287Mb)
    Date
    2022
    Author
    Inacio, Jordan V.
    Schwarzenberg, Peter
    Yoon, Richard S.
    Kantzos, Andrew
    Malige, Ajith
    Nwachuku, Chinenye O.
    Dailey, Hannah L.
    Metadata
    Show full item record
    Abstract
    The mechanics of distal femur fracture fixation has been widely studied in bench tests that employ a variety of approaches for holding and constraining femurs to apply loads. No standard test methods have been adopted for these tests and the impact of test setup on inferred construct mechanics has not been reported. Accordingly, the purpose of this study was to use finite element models to compare the mechanical performance of a supracondylar osteotomy with lateral plating under conditions that replicate several common bench test methods. A literature review was used to define a parameterized virtual model of a plated distal femur osteotomy in axial compression loading with four boundary condition sets ranging from minimally to highly constrained. Axial stiffness, fracture gap closure, and transverse motion at the fracture line were recorded for a range of applied loads and bridge spans. The results showed that construct mechanical performance was highly sensitive to boundary conditions imposed by the mechanical test fixtures. Increasing the degrees of constraint, for example, by potting and rigidly clamping one or more ends of the specimen, caused up to a 25× increase in axial stiffness of the construct. Transverse motion and gap closure at the fracture line, which is an important driver of interfragmentary strain, was also largely influenced by the constraint test setup. These results suggest that caution should be used when comparing reported results between bench tests that use different fixtures and that standardization of testing methods is needed in this field.
    URI
    http://yetl.yabesh.ir/yetl1/handle/yetl/4284026
    Collections
    • Journal of Biomechanical Engineering

    DSpace software copyright © 2002-2015  DuraSpace
    Contact Us | Send Feedback
    Theme by 
    @mire NV
     

     

    Browse

    All of DSpaceCommunities & CollectionsBy Issue DateAuthorsTitlesSubjectsThis CollectionBy Issue DateAuthorsTitlesSubjects

    My Account

    LoginRegister

    Statistics

    View Usage Statistics

    DSpace software copyright © 2002-2015  DuraSpace
    Contact Us | Send Feedback
    Theme by 
    @mire NV