Show simple item record

dc.contributor.authorCastro, Miguel Nobre
dc.contributor.authorRahman, Tariq
dc.contributor.authorNicholson, Kristen Faith
dc.contributor.authorRasmussen, John
dc.contributor.authorBai, Shaoping
dc.contributor.authorAndersen, Michael Skipper
dc.date.accessioned2022-02-04T22:10:53Z
dc.date.available2022-02-04T22:10:53Z
dc.date.copyright6/26/2020 12:00:00 AM
dc.date.issued2020
dc.identifier.issn1932-6181
dc.identifier.otherjesmdt_003_03_031005.pdf
dc.identifier.urihttp://yetl.yabesh.ir/yetl1/handle/yetl/4275039
dc.description.abstractPassive arm-assistive devices play an important role in the rehabilitation of patients with neuromuscular disorders or injuries by overcoming their motor deficit. Routine human activities such as feeding are not possible without the aid provided by one of these devices or by a caregiver. In this study, a body-powered assistive device was designed for feeding purposes using a compact spherical scissors mechanism and zero-free-length (ZFL) springs (rubber bands) to leverage the patient's residual biceps and healthy triceps function. This partially balanced and lightweight orthosis was also projected to accommodate the spring attachment points closer to the elbow joint center. The performance of the prototype was evaluated on a young adult with bilateral amyoplasia of the biceps due to arthrogryposis who could not initially reach the superior anterior aspect of the close-to-torso region of the reachable three-dimensional (3D) workspace (RWS). That was accomplished by measuring the anatomical RWS of the patient before and while wearing the device. The results show that the patient, with the assistance provided by the device, was able to attain positions in the frontal close-to-torso region of the body that included reaching her mouth, thus enabling independent feeding.
dc.publisherThe American Society of Mechanical Engineers (ASME)
dc.titleA Case Study on Designing a Passive Feeding-Assistive Orthosis for Arthrogryposis
dc.typeJournal Paper
dc.journal.volume14
dc.journal.issue3
dc.journal.titleJournal of Medical Devices
dc.identifier.doi10.1115/1.4047462
dc.journal.fristpage034502-1
dc.journal.lastpage034502-8
dc.page8
dc.treeJournal of Medical Devices:;2020:;volume( 014 ):;issue: 003
dc.contenttypeFulltext


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record