Show simple item record

dc.contributor.authorChen, Tianhao
dc.contributor.authorLancaster, Michael
dc.contributor.authorLin, Dawn S. Y.
dc.contributor.authorDoyle, Matthew G.
dc.contributor.authorForbes, Thomas L.
dc.contributor.authorAmon, Cristina H.
dc.date.accessioned2019-09-18T09:07:10Z
dc.date.available2019-09-18T09:07:10Z
dc.date.copyright4/11/2019 12:00:00 AM
dc.date.issued2019
dc.identifier.issn1932-6181
dc.identifier.othermed_013_02_021008
dc.identifier.urihttp://yetl.yabesh.ir/yetl1/handle/yetl/4259078
dc.description.abstractStent grafts are medical devices used to treat abdominal aortic aneurysms (AAAs) in endovascular aneurysm repair (EVAR). Computational and experimental models have been developed to study stent graft delivery and deployment during EVAR; however, frictional properties have not been taken into account in most previous studies. The objective of this study was to determine the coefficients of friction of three commercially available stent grafts (Cook Zenith, Medtronic Endurant, and Vascutek Anaconda), their delivery sheaths, a porcine aorta, and two mock arterial materials. Stent grafts were obtained and separated into stents, graft fabric, and sheaths. Using a custom-made friction measurement apparatus, the coefficients of friction were measured between five material pairs: (i) the stents and inner surface of the sheath, (ii) the graft fabric and inner surface of the sheath, (iii) the outer surface of the sheath and a porcine aorta, (iv) the outer surface of the sheath and three different polyvinyl alcohol (PVA) cryogels, and (v) the outer surface of the sheath and a polydimethylsiloxane (PDMS) sheet. The results show that the coefficients of friction between the graft fabric and the sheath were higher than those between the stents and the sheath. The PVA cryogels showed more comparable frictional properties to the porcine aorta than did the PDMS sheet, suggesting that PVA cryogels provide a more accurate approximation for the in vivo frictional properties. These results can be used to improve the accuracy of computational models for stent graft delivery and deployment and to select appropriate materials for vascular phantoms.
dc.publisherAmerican Society of Mechanical Engineers (ASME)
dc.titleMeasurement of Frictional Properties of Aortic Stent Grafts and Their Delivery Systems
dc.typeJournal Paper
dc.journal.volume13
dc.journal.issue2
dc.journal.titleJournal of Medical Devices
dc.identifier.doi10.1115/1.4043292
dc.journal.fristpage21008
dc.journal.lastpage021008-9
dc.treeJournal of Medical Devices:;2019:;volume( 013 ):;issue: 002
dc.contenttypeFulltext


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record