dc.description.abstractNeedle biopsy is a routine medical procedure for examining tissue or biofluids for the presence of disease using standard methods of pathology. The finite element analysis (FEA) methodology can provide guidance for optimizing the geometric parameters. The needle biopsy is simulated and analyzed while varying the needle angle, the aperture size and the slice-push ratio k. The results indicate that tissue reaction force in the axial direction of needle gradually decreases, and the stress and strain are more concentrated at the tip of needle with the increases of tip angle; the tissue reaction force decreases, and the torque increases while the slice-push ratio increases; and higher slice–push ratio can increase the peak stress concentration on the cutting edge and deformation of tissue; in the process of core needle cutting, increasing slice–push ratio can reduce the tissue reaction force significantly. While the aperture on distal wall of outer cannula becomes wider, the tissue reaction force increases significantly, and the cutting process will be more unstable. The results have the potential to provide important insight for improving the needle biopsy design process.
dc.publisherThe American Society of Mechanical Engineers (ASME)
dc.titleUsing Simulation to Help Specify Design Parameters for Vacuum-Assisted Needle Biopsy Systems
dc.typeJournal Paper
dc.journal.volume13
dc.journal.issue1
dc.journal.titleJournal of Medical Devices
dc.identifier.doi10.1115/1.4041487
dc.journal.fristpage14502
dc.journal.lastpage014502-7
dc.treeJournal of Medical Devices:;2019:;volume( 013 ):;issue: 001