    Improved Method for the ASTM Platelet and Leukocyte Assay: Use of Minimal Heparinization in a Screening Test for Hemocompatibility of Blood-Contacting Medical Devices

    Thumbnail
    View/Open
    med_013_01_011004.pdf (1.821Mb)
    Date
    2019
    Author
    Schatz, Tim
    Howard, Sarah
    Porter, Deanna
    Grove, Kent
    Smith, Mark E.
    Chen, Yan
    Metadata
    Show full item record
    Abstract
    Most blood-contacting medical devices must be assessed for potential thrombogenicity prior to regulatory approval. A common assay for screening and qualifying devices involves monitoring the reduction of platelet and leukocyte (P&L) counts in whole blood exposed to the device. We have validated an improved method for assessing a device's effect on platelet activation and surface adhesion, offering significant improvement over the current ASTM F2888-13 method, which uses blood fully anticoagulated by acidified citrate (known to significantly inhibit platelet responsiveness). Our method uses minimal heparinization (final concentration 1 IU/mL) to optimize the response to commonly used control materials: latex, black rubber, and high-density polyethylene (HDPE). We also have shown the assay's capacity to appropriately assess a legally marketed comparator device (LMCD) with a documented clinical history. The test materials were prepared for incubation and allowed to remain in contact with the citrated or heparinized blood for ∼1 h at 37 °C. A complete blood count was performed prior to exposure, and at the end of the incubation period, reductions in P&L counts were recorded. Results from citrate-anticoagulated assay showed only a marginal response to the positive control, black rubber. Using heparinized blood, the assay generated a robust response to the positive controls, the “intermediate scoring” controls, and also assessed a legally marketed and approved device as clearly nonthrombogenic. This modification adds robustness and sensitivity to this quick and inexpensive thrombogenicity assay and should be incorporated into the next ASTM standards.
    URI
    http://yetl.yabesh.ir/yetl1/handle/yetl/4256416
    Collections
    • Journal of Medical Devices

    DSpace software copyright © 2002-2015  DuraSpace
    Contact Us | Send Feedback
    Theme by 
    @mire NV
     

     

    Browse

    All of DSpaceCommunities & CollectionsBy Issue DateAuthorsTitlesSubjectsThis CollectionBy Issue DateAuthorsTitlesSubjects

    My Account

    LoginRegister

    Statistics

    View Usage Statistics

    DSpace software copyright © 2002-2015  DuraSpace
    Contact Us | Send Feedback
    Theme by 
    @mire NV