dc.contributor.authorOkkalidis, Nikiforos
dc.contributor.authorChatzigeorgiou, Chrysoula
dc.contributor.authorOkkalides, Demetrios
dc.date.accessioned2019-03-17T10:13:34Z
dc.date.available2019-03-17T10:13:34Z
dc.date.copyright11/7/2017 12:00:00 AM
dc.date.issued2019
dc.identifier.issn2572-7958
dc.identifier.otherjesmdt_001_01_011003.pdf
dc.identifier.urihttp://yetl.yabesh.ir/yetl1/handle/yetl/4256001
dc.description.abstractA couple of fused deposition modeling (FDM) three-dimensional (3D) printers using variable infill density patterns were employed to simulate human muscle, fat, and lung tissue as it is represented by Hounsfield units (HUs) in computer tomography (CT) scans. Eleven different commercial plastic filaments were assessed by measuring their mean HU on CT images of small cubes printed with different patterns. The HU values were proportional to the mean effective density of the cubes. Polylactic acid (PLA) filaments were chosen. They had good printing characteristics and acceptable HU. Such filaments obtained from two different vendors were then tested by printing two sets of cubes comprising 10 and 6 cubes with 100% to 20% and 100% to 50% infill densities, respectively. They were printed with different printing patterns named “Regular” and “Bricks,” respectively. It was found that the HU values measured on the CT images of the 3D-printed cubes were proportional to the infill density with slight differences between vendors and printers. The Regular pattern with infill densities of about 30%, 90%, and 100% were found to produce HUs equivalent to lung, fat, and muscle. This was confirmed with histograms of the respective region of interest (ROI). The assessment of popular 3D-printing materials resulted in the choice of PLA, which together with the proposed technique was found suitable for the adequate simulation of the muscle, fat, and lung HU in printed patient-specific phantoms.
dc.publisherThe American Society of Mechanical Engineers (ASME)
dc.titleAssessment of 11 Available Materials With Custom Three-Dimensional-Printing Patterns for the Simulation of Muscle, Fat, and Lung Hounsfield Units in Patient-Specific Phantoms
dc.typeJournal Paper
dc.journal.volume1
dc.journal.issue1
dc.journal.titleJournal of Engineering and Science in Medical Diagnostics and Therapy
dc.identifier.doi10.1115/1.4038228
dc.journal.fristpage11003
dc.journal.lastpage011003-7
dc.treeJournal of Engineering and Science in Medical Diagnostics and Therapy:;2019:;volume( 001 ):;issue: 001