Show simple item record

dc.contributor.authorSoenksen, Luis R.
dc.contributor.authorMartínez-Corona, David A.
dc.contributor.authorIñiguez de Gante, Sofía
dc.contributor.authorPhabmixay, Pierre S.
dc.contributor.authorMarongiu Maggi, Mauricio J.
dc.date.accessioned2019-02-28T11:04:58Z
dc.date.available2019-02-28T11:04:58Z
dc.date.copyright1/30/2018 12:00:00 AM
dc.date.issued2018
dc.identifier.issn1932-6181
dc.identifier.othermed_012_01_011009.pdf
dc.identifier.urihttp://yetl.yabesh.ir/yetl1/handle/yetl/4252482
dc.description.abstractThe shelf life of point-of-care and rapid diagnostic tests (POC-RDTs) is commonly compromised by abrupt temperature changes during storage, transportation, and use. This situation is especially relevant in tropical regions and resource-constrained settings where cold chain may be unreliable. Here, we report the use of novel and low-cost passive thermal shield (TS) made from laminated phase change material (PCM) to reduce thermal overload in POC-RDTs. Validation of the proposed design was done through numerical simulation and testing of an octadecane shield prototype in contact with a lateral flow immunoassay. The use of our TS design provided 30–45 min delay in thermal equilibration under constant and oscillating heat load challenges resembling those of field use. The addition of a thin PCM protection layer to POC-RDTs can be a cost-effective, scalable, and reliable solution to provide additional thermal stability to these devices.
dc.publisherThe American Society of Mechanical Engineers (ASME)
dc.titleLow-Cost Thermal Shield for Rapid Diagnostic Tests Using Phase Change Materials
dc.typeJournal Paper
dc.journal.volume12
dc.journal.issue1
dc.journal.titleJournal of Medical Devices
dc.identifier.doi10.1115/1.4038898
dc.journal.fristpage11009
dc.journal.lastpage011009-12
dc.treeJournal of Medical Devices:;2018:;volume( 012 ):;issue: 001
dc.contenttypeFulltext


Files in this item

Thumbnail
This item appears in the following Collection(s)

Show simple item record