    Numerical and Experimental Simulations of the Wireless Energy Transmission and Harvesting by a Camera Pill

    Thumbnail
    View/Open
    med_012_02_021002.pdf (1.193Mb)
    Date
    2018
    Author
    Shumbayawonda, Elizabeth
    Salifu, Ali A.
    Lekakou, Constantina
    Cosmas, John P.
    Metadata
    Show full item record
    Abstract
    This paper investigates the energy transmitted to and harvested by a camera pill traveling along the gastrointestinal (GI) tract. It focuses on the transmitted electromagnetic (EM) energy in the frequency range of 0.18 to 2450 MHz and compares it to the mechanical energy due to the motion of the pill and the force exerted from the intestine in its peristalsis onto the pill, and the electrochemical energy due to the change of pH along the path of the pill. A comprehensive multilayer EM power transmission model is constructed and implemented in a numerical code, including power attenuation through each layer and multireflections at material interfaces. Computer simulations of EM power transmission through a multilayer abdomen to a pill traveling in the intestine are presented for the human abdominal cavity as well as phantom organs and phantom environments, coupled with corresponding experimental studies using these phantom components and environments. Two types of phantom abdomen are investigated: a ballistic gel and a multilayer duck breast. Phantom small intestine involves gelatin gel layers with embedded phantom chyme. Due to limitations related to the energy safety limit of skin exposure and energy losses in the transmission through the abdomen and intestines, inductive range frequencies are recommended which may yield energy harvesting of 10–50 mWh during 8 h of pill journey, complemented by about 10 mWh of mechanical energy and 10 mWh of electrochemical energy harvesting, in addition to about 330 mWh typically stored in the coin batteries of a camera pill.
    URI
    http://yetl.yabesh.ir/yetl1/handle/yetl/4252480
    Collections
    • Journal of Medical Devices

    DSpace software copyright © 2002-2015  DuraSpace
    Contact Us | Send Feedback
    Theme by 
    @mire NV
     

     

    Browse

    All of DSpaceCommunities & CollectionsBy Issue DateAuthorsTitlesSubjectsThis CollectionBy Issue DateAuthorsTitlesSubjects

    My Account

    LoginRegister

    Statistics

    View Usage Statistics

    DSpace software copyright © 2002-2015  DuraSpace
    Contact Us | Send Feedback
    Theme by 
    @mire NV