    A Novel, Needle-Array Dry-Electrode With Stainless Steel Micro-Tips, for Electroencephalography Monitoring

    Thumbnail
    View/Open
    med_012_04_041001.pdf (2.792Mb)
    Date
    2018
    Author
    Radhakrishnan, J. K.
    Nithila, S.
    Kartik, S. N.
    Bhuvana, T.
    Kulkarni, G. U.
    Singh, U. K.
    Metadata
    Show full item record
    Abstract
    A novel, needle array dry electrode consisting of 10 × 10 array of stainless steel (SS) Microtips was developed for electroencephalography (EEG) monitoring. The developed dry electrode uses commercially available, inexpensive, SS acupuncture needles certified for invasive use, to collect the EEG signal. The microtips of the acupuncture needles project out of a flat Teflon base by approximately 150 μm. Mechanical failure analysis was carried out, with theoretical calculations for individual needles and experimental measurements with a universal testing machine (UTM). The theoretically calculated critical load for failure for individual needle was 0.88 N, while the UTM measurements show the failure occurring at 0.95 N; this difference is probably due to the simplified assumptions used in calculations. The UTM measurements of the individual needle applied against a Silicone elastomer reveal that the force required for the penetration of the needle of the electrode into skin maybe as low as 0.01 N. Needle array insertion into silicone elastomer sheet and its optical inspection was carried out to assess the ability of the microneedles to penetrate the skin. The impedance of the electrode, measured in three electrode configuration in 0.9% NaCl solution, was approximately 6.8KΩ at 20 Hz, which is sufficiently low to fulfill the requirements of biopotential measurement. The construction and characteristics of the developed needle array dry electrode show that they are suitable for penetrating the stratum corneum of the skin and acquire the EEG signal directly from the interstitial fluidic layer underneath. The construction of the electrode and its mechanical and electrical characteristics show that it is a promising dry electrode for long duration EEG Monitoring.
    URI
    http://yetl.yabesh.ir/yetl1/handle/yetl/4252471
    Collections
    • Journal of Medical Devices

    DSpace software copyright © 2002-2015  DuraSpace
    Contact Us | Send Feedback
    Theme by 
    @mire NV
     

     

    Browse

    All of DSpaceCommunities & CollectionsBy Issue DateAuthorsTitlesSubjectsThis CollectionBy Issue DateAuthorsTitlesSubjects

    My Account

    LoginRegister

    Statistics

    View Usage Statistics

    DSpace software copyright © 2002-2015  DuraSpace
    Contact Us | Send Feedback
    Theme by 
    @mire NV