Show simple item record

dc.contributor.authorNg, Cailin
dc.contributor.authorLiang, Wenyu
dc.contributor.authorGan, Chee Wee
dc.contributor.authorLim, Hsueh Yee
dc.contributor.authorTan, Kok Kiong
dc.date.accessioned2019-02-28T11:04:53Z
dc.date.available2019-02-28T11:04:53Z
dc.date.copyright7/13/2018 12:00:00 AM
dc.date.issued2018
dc.identifier.issn1932-6181
dc.identifier.othermed_012_03_031004.pdf
dc.identifier.urihttp://yetl.yabesh.ir/yetl1/handle/yetl/4252467
dc.description.abstractAn automated surgical device, the ventilator tube applicator (VTA), enables a grommet insertion surgery for patients with otitis media with effusion (OME) to be completed in a short time automatically and precisely, eliminating the use of general anesthesia (GA) typically required in such procedure. However, its current design limits the usefulness of the device as it is restricted by the properties of the tympanic membrane (TM), such as angle, thickness, and strength. Therefore, a novel design was conceptualized and the insertion control algorithm was improved to overcome the current challenges of the VTA. This innovative cover-cutter instrument design allows three-dimensional (3D) motion on an oblique surface using a single axis actuator. Experimental results on mock membranes showed great improvements in terms of robustness and success rate. The new design allowed the procedure to be performed on wider range of TM angles and hence increased the effectiveness of VTA. Grommet insertion force was reduced by an average of 66%, and the overall peak force reduced by an average of 14%. Finite element (FE) analysis on a cadaveric TM model further validated the usefulness of the cover-cutter instrument, and showed some interesting insights in the grommet insertion process.
dc.publisherThe American Society of Mechanical Engineers (ASME)
dc.titleNovel Design and Validation of a Micro Instrument in an Ear Grommet Insertion Device
dc.typeJournal Paper
dc.journal.volume12
dc.journal.issue3
dc.journal.titleJournal of Medical Devices
dc.identifier.doi10.1115/1.4040185
dc.journal.fristpage31004
dc.journal.lastpage031004-12
dc.treeJournal of Medical Devices:;2018:;volume( 012 ):;issue: 003
dc.contenttypeFulltext


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record