Show simple item record

dc.contributor.authorAharonson, Vered
dc.contributor.authorSchlesinger, Ilana
dc.contributor.authorMcDonald, Andre M.
dc.contributor.authorDubowsky, Steven
dc.contributor.authorKorczyn, Amos D.
dc.date.accessioned2019-02-28T11:04:43Z
dc.date.available2019-02-28T11:04:43Z
dc.date.copyright2/12/2018 12:00:00 AM
dc.date.issued2018
dc.identifier.issn1932-6181
dc.identifier.othermed_012_01_011012.pdf
dc.identifier.urihttp://yetl.yabesh.ir/yetl1/handle/yetl/4252437
dc.description.abstractWe present personal aid for mobility and monitoring (PAMM II), an instrumented walker for Parkinson's disease (PD) patients' gait monitoring. The objective of the walker is to aid in the diagnosis and monitoring of PD progression as well as the effects of clinical treatment and rehabilitation. In contrast to existing devices, the walker is a low-cost solution that is simple to operate and maintain, requiring no adjustments, special usage instructions, or infrastructure. This preliminary study reports on the efficiency, reliability, and accuracy of PAMM II when used to evaluate 22 PD patients and 20 control individuals. All subjects walked two prescribed paths while pushing the walker, and their kinematic motion signals were automatically collected by the walker. Feature derivation from the walker's signals was followed by combinations of two classical feature selection methods and two learning algorithms, with the objective of discriminating PD patients from control subjects. Sensitivity and specificity scores of 91% and 95% were achieved for the first walking protocol, whereas discrimination over the second walking protocol produced sensitivity and specificity scores of 96% and 100%. These preliminary results provide insight as to the usefulness of PAMM II and its data processing algorithms for the assessment of PD patients' condition.
dc.publisherThe American Society of Mechanical Engineers (ASME)
dc.titleA Practical Measurement of Parkinson's Patients Gait Using Simple Walker-Based Motion Sensing and Data Analysis
dc.typeJournal Paper
dc.journal.volume12
dc.journal.issue1
dc.journal.titleJournal of Medical Devices
dc.identifier.doi10.1115/1.4038810
dc.journal.fristpage11012
dc.journal.lastpage011012-8
dc.treeJournal of Medical Devices:;2018:;volume( 012 ):;issue: 001
dc.contenttypeFulltext


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record