    Surface and Thermal Characteristics of Single-Use Electrosurgical Pencils After Clinical Reuse and In-Hospital Reprocessing

    Thumbnail
    View/Open
    med_011_04_041010.pdf (3.369Mb)
    Date
    2017
    Author
    Tessarolo
    Francesco;Torres
    Sebastian;Ballesteros
    Luis Miguel;Rigoni
    Marta;Piccoli
    Federico;Caola
    Iole;Caciagli
    Patrizio;Montoya
    Yesid;Nollo
    Giandomenico
    Metadata
    Show full item record
    Abstract
    Safety and efficacy issues are associated with reprocessing of single-use electrosurgical pencils (EPs), requiring methods for assessing the reprocessing protocol before clinical reuse. This study aimed at monitoring the surface characteristics of single-use EPs subjected to multiple clinical use and in-hospital reprocessing. A total of 24 single-use-labeled EPs were divided in five test groups and one control group. The test groups were subjected to a different number of clinical uses, ranging from one to five. A multitechnique approach based on optical stereomicroscopy (OM), scanning electron microscopy (SEM), energy dispersive X-ray spectroscopy (EDXS), differential scanning calorimetry (DSC), and thermogravimetric analysis (TGA) was applied. The silicon coating of the tip was significantly reduced, and foreign bodies were occasionally found on reprocessed EPs. The amount of biological debris and chemical residuals increased with the number of reprocessing cycles in critical areas. The degradation temperature of the EP handle polymer showed a progressive significant reduction. Cable cord showed no modification after reprocessing. EP tip could undergo major surface modifications that can affect functionality. The efficacy of the reprocessing protocol in removing debris from the EP handle should be carefully assessed. Surface and thermal characteristics have to be considered for validating a reprocessing protocol of single-use labeled EP.
    URI
    http://138.201.223.254:8080/yetl1/handle/yetl/4242824
    Collections
    • Journal of Medical Devices

    DSpace software copyright © 2002-2015  DuraSpace
    Contact Us | Send Feedback
    Theme by 
    @mire NV
     

     

    Browse

    All of DSpaceCommunities & CollectionsBy Issue DateAuthorsTitlesSubjectsThis CollectionBy Issue DateAuthorsTitlesSubjects

    My Account

    LoginRegister

    Statistics

    View Usage Statistics

    DSpace software copyright © 2002-2015  DuraSpace
    Contact Us | Send Feedback
    Theme by 
    @mire NV