    A Human Thermoregulation Simulator for Calibrating Water-Perfused Cooling Pad Systems for Therapeutic Hypothermia

    Thumbnail
    View/Open
    med_011_03_034506.pdf (3.127Mb)
    Date
    2017
    Author
    Chacko, Priya S. E.
    Seifi, Ali
    Diller, Kenneth R.
    Metadata
    Show full item record
    Abstract
    The induction of a mild reduction in body core temperature has been demonstrated to provide neuroprotection for patients who have suffered a medical event resulting in ischemia to the brain or vital organs. Temperatures in the range of 32–34 °C provide the required level of protection and can be produced and maintained by diverse means for periods of days. Rewarming from hypothermia must be conducted slowly to avoid serious adverse consequences and usually is performed under control of the thermal therapeutic device based on a closed-loop feedback strategy based on the patient's core temperature. Given the sensitivity and criticality of this process, it is important that the device control system be able to interact with the human thermoregulation system, which itself is highly nonlinear. The therapeutic hypothermia device must be calibrated periodically to ensure that its performance is accurate and safe for the patient. In general, calibration processes are conducted with the hypothermia device operating on a passive thermal mass that behaves much differently than a living human. This project has developed and demonstrated an active human thermoregulation simulator (HTRS) that embodies major governing thermal functions such as central metabolism, tissue conduction, and convective transport between the core and the skin surface via the flow of blood and that replicates primary dimensions of the torso. When operated at physiological values for metabolism and cardiac output, the temperature gradients created across the body layers and the heat exchange with both an air environment and a clinical water-circulating cooling pad system match that which would occur in a living body. Approximately two-thirds of the heat flow between the core and surface is via convection rather than conduction, highlighting the importance of including the contribution of blood circulation to human thermoregulation in a device designed to calibrate the functioning of a therapeutic hypothermia system. The thermoregulation simulator functions as anticipated for a typical living patient during both body cooling and warming processes. This human thermoregulatory surrogate can be used to calibrate the thermal function of water-perfused cooling pads for a hypothermic temperature management system during both static and transient operation.
    URI
    http://138.201.223.254:8080/yetl1/handle/yetl/4235245
    Collections
    • Journal of Medical Devices

    DSpace software copyright © 2002-2015  DuraSpace
    Contact Us | Send Feedback
    Theme by 
    @mire NV
     

     

    Browse

    All of DSpaceCommunities & CollectionsBy Issue DateAuthorsTitlesSubjectsThis CollectionBy Issue DateAuthorsTitlesSubjects

    My Account

    LoginRegister

    Statistics

    View Usage Statistics

    DSpace software copyright © 2002-2015  DuraSpace
    Contact Us | Send Feedback
    Theme by 
    @mire NV