    RFID-Based Real-Time Navigation for Interventional Magnetic Resonance Imaging: Development and Evaluation of a Novel Tracking System

    Thumbnail
    View/Open
    med_011_03_031007.pdf (2.057Mb)
    Date
    2017
    Author
    Güttler, Felix
    Heinrich, Andreas
    Krauß, Peter
    Guntermann, Jonathan
    de Bucourt, Maximilian
    Teichgräber, Ulf
    Metadata
    Show full item record
    Abstract
    The purpose of this study was to evaluate the suitability of a novel radio-frequency identification (RFID)-based tracking system for intraoperative magnetic resonance imaging (MRI). A RFID tracking system was modified to fulfill MRI-compatibility and tested according to ASTM and NEMA. The influence of the RFID tracking system on MRI was analyzed in a phantom study using a half-Fourier acquisition single-shot turbospin echo (HASTE) and true fast imaging with steady-state precession sequence (TrueFISP) sequence. The RFID antenna was gradually moved closer to the isocenter of the MR scanner from 90 to 210 cm to investigate the influence of the distance. Furthermore, the RF was gradually changed between 865 and 869 MHz for a distance of 90 cm, 150 cm, and 210 cm to the isocenter of the magnet to investigate the influence of the frequency. The specific spatial resolution was measured with and without a permanent line of sight (LOS). After the modification of the reader, no significant change of the signal-to-noise ratio (SNR) could be observed with increasing distance of the RFID tracking system to the isocenter of the MR scanner. Also, different radio frequencies of the RFID tracking system did not influence the SNR of the MR-images significantly. The specific spatial resolution deviated on average by 8.97 ± 7.33 mm with LOS and 11.23 ± 12.03 mm without LOS from the reference system. The RFID tracking system had no relevant influence on the MR-image quality. RFID tracking solved the LOS problem. However, the spatial accuracy of the RFID tracking system has to be improved for medical usage.
    URI
    http://138.201.223.254:8080/yetl1/handle/yetl/4235232
    Collections
    • Journal of Medical Devices

    DSpace software copyright © 2002-2015  DuraSpace
    Contact Us | Send Feedback
    Theme by 
    @mire NV
     

     

    Browse

    All of DSpaceCommunities & CollectionsBy Issue DateAuthorsTitlesSubjectsThis CollectionBy Issue DateAuthorsTitlesSubjects

    My Account

    LoginRegister

    Statistics

    View Usage Statistics

    DSpace software copyright © 2002-2015  DuraSpace
    Contact Us | Send Feedback
    Theme by 
    @mire NV