• Login
    View Item 
    •   DSpace Home
    • ASME
    • Journal of Medical Devices
    • View Item
    •   DSpace Home
    • ASME
    • Journal of Medical Devices
    • View Item
    JavaScript is disabled for your browser. Some features of this site may not work without it.

    Continuous, Real-Time, Noninvasive Hemodynamic Cardiac Doppler Monitoring With a Novel Hands-Free Device: A Feasibility Study Compared to Standard Echo

    Thumbnail
    View/Open
    med_011_03_031001.pdf (612.8Kb)
    Date
    2017
    Author
    Lancaster, Gilead I.
    Hay, Ilan
    Eldar, Michael
    Metadata
    Show full item record
    Abstract
    This study sought to determine the feasibility of using noninvasive cardiac hemodynamics (NICHE), a new noninvasive Doppler-based device, to monitor real-time, simultaneous tissue and blood-flow Doppler measurements in a clinical setting, and to obtain preliminary performance data compared to a commercially available system. Doppler-based measurements have been shown to correlate well with invasive hemodynamic data and diastolic function, but their use in clinical applications has been limited by various technical issues. The NICHE device was developed to obtain simultaneous tissue and blood-flow Doppler measurements automatically, in real-time and in a hands-free manner. Thirty participants (ten normal volunteers and 20 patients in a cardiac rehab program) underwent standard echocardiographic/Doppler studies followed immediately by NICHE monitoring. Early diastolic transmitral blood-flow velocity (E) and tissue Doppler myocardial wall velocity during early relaxation (E′) were acquired using a standard echo device; and E/E′ was derived post hoc. NICHE measurements included E, E′, and directly measured instantaneous E/E′. NICHE was successfully used in 28 participants. Measurements of ENICHE ranged from 40 cm/s to over 120 cm/s and correlated well with Eecho (R = 0.93). ENICHE′ ranged from 2 to 23 cm/s and correlated well with the averaged Eecho′ (R = 0.91). Directly measured E/ENICHE′ ratios ranged from 3 to 23 and correlated well with derived E/Eecho′ (R = 0.91). The NICHE device can monitor patients in a hands-free manner and can supply real-time Doppler derived measurements of hemodynamic parameters and diastolic function that correlate well with measurements from standard echo devices.
    URI
    http://138.201.223.254:8080/yetl1/handle/yetl/4235225
    Collections
    • Journal of Medical Devices

    DSpace software copyright © 2002-2015  DuraSpace
    Contact Us | Send Feedback
    Theme by 
    @mire NV
     

     

    Browse

    All of DSpaceCommunities & CollectionsBy Issue DateAuthorsTitlesSubjectsThis CollectionBy Issue DateAuthorsTitlesSubjects

    My Account

    LoginRegister

    Statistics

    View Usage Statistics

    DSpace software copyright © 2002-2015  DuraSpace
    Contact Us | Send Feedback
    Theme by 
    @mire NV