    Automated Real-Time Processing of Single Lead Electrocardiogram for Simultaneous Heart Rate and Respiratory Rate Monitoring

    Thumbnail
    View/Open
    med_011_02_024502.pdf (538.8Kb)
    Date
    2017
    Author
    Dutta, Disha N.
    Das, Reshmi
    Pal, Saurabh
    Metadata
    Show full item record
    Abstract
    In this article, the design and development of a real-time heart rate (HR) and respiratory rate (RR) monitoring device is reported. The proposed device is designed to impose minimum data acquisition hazards on the subject. In standard bedside monitors, HR and RR are derived from electrocardiogram (ECG) and respiration signals, respectively, and different electrodes are required for capturing the 12-lead ECG and respiration via a chest belt, which is cumbersome for patients and healthcare providers. Respiration signal has an impact on ECG due to anatomical proximity of the heart and lung, and ECG is modulated by respiration, a phenomenon known as respiratory sinus arrhythmia (RSA). In the proposed method, the ECG signal is acquired using clip electrodes at the wrists and the respiration signal is extracted from the ECG using an Arduino Uno microcontroller-based real-time processing of ECG. RR is then derived from ECG-derived respiration (EDR). The prototype is tested on healthy subjects and compared to measurements taken using a standard MP45 data acquisition device associated with a Biopac Student Lab (BSL). A mean percentage error of 5.54 ± 8.48% was observed under normal breathing conditions and an error of −3.41 ± 3.27% was observed for a single subject tested under a variety of breathing conditions, such as resting, stair-climbing, and paced breathing. The proposed algorithm can also be used in combination with standard ECG monitoring systems to measure HR and RR, without any data acquisition hazard to the subject.
    URI
    http://138.201.223.254:8080/yetl1/handle/yetl/4235223
    Collections
    • Journal of Medical Devices

    DSpace software copyright © 2002-2015  DuraSpace
    Contact Us | Send Feedback
    Theme by 
    @mire NV
     

     

    Browse

    All of DSpaceCommunities & CollectionsBy Issue DateAuthorsTitlesSubjectsThis CollectionBy Issue DateAuthorsTitlesSubjects

    My Account

    LoginRegister

    Statistics

    View Usage Statistics

    DSpace software copyright © 2002-2015  DuraSpace
    Contact Us | Send Feedback
    Theme by 
    @mire NV