Show simple item record

dc.contributor.authorPike, Julian M.
dc.date.accessioned2017-06-09T16:33:39Z
dc.date.available2017-06-09T16:33:39Z
dc.date.copyright1984/06/01
dc.date.issued1984
dc.identifier.issn0739-0572
dc.identifier.otherams-70.pdf
dc.identifier.urihttp://onlinelibrary.yabesh.ir/handle/yetl/4211733
dc.description.abstractField instruments require calibration against standards that introduce-insignificant errors of their own in the process. This paper discusses a pressure and temperature calibration facility, analyzes the error levels attained in its routine use and shows that the accuracies are both reasonable and adequate. The results are given in terms of uncertainty; values of 0.025 kPa for pressure and 0.040 K for temperature were shown to meet the needs for calibration of research quality sensors.
dc.publisherAmerican Meteorological Society
dc.titleRealistic Uncertainties in Pressure and Temperature Calibration Reference Values
dc.typeJournal Paper
dc.journal.volume1
dc.journal.issue2
dc.journal.titleJournal of Atmospheric and Oceanic Technology
dc.identifier.doi10.1175/1520-0426(1984)001<0115:RUIPAT>2.0.CO;2
dc.journal.fristpage115
dc.journal.lastpage119
dc.treeJournal of Atmospheric and Oceanic Technology:;1984:;volume( 001 ):;issue: 002