Show simple item record

dc.contributor.authorArun, Anupama
dc.contributor.authorGaston, Brandon
dc.contributor.authorKwiat, Dillon
dc.contributor.authorImamura
dc.contributor.authorFechter, Richard
dc.contributor.authorChen, Shijie
dc.contributor.authorJiang, Hao
dc.contributor.authorHarrison, Michael
dc.contributor.authorRoy, Shuvo
dc.date.accessioned2017-05-09T01:31:41Z
dc.date.available2017-05-09T01:31:41Z
dc.date.issued2016
dc.identifier.issn1932-6181
dc.identifier.othermed_010_01_011008.pdf
dc.identifier.urihttp://yetl.yabesh.ir/yetl/handle/yetl/161987
dc.description.abstractMagnetic compression based anastomoses use magnetic force to necrose tissue between two magnets to create an anastomosis. Nickelplated neodymium–iron–boron magnets are used in our study. The compression pressure between the magnets depends on the distance between the magnets, which is determined by the thickness of the compressed tissue and depends on bowel wall thickness and elasticity. It is critical to know the distance between the magnets once the tissue is compressed because the magnets must be within a critical distance of each other in order to create enough compressive force to necrose the tissue. We have developed an inductance sensor to detect the distance (tissue thickness) between the two magnets after the surgeon has deployed them. Inductance sensing is a contactless sensing method that enables precise shortrange detection of conducting surfaces. The inductor coil mounted on one magnet detects the second magnet by measuring the change in inductance due to eddy current induced on the nickelplated surface of the second magnet. The change in the inductance is proportional to the change in distance between the magnets. The sensor was first calibrated by using polycarbonate sheets to simulate the intestine tissue. We are able to detect up to 6 mm of spacing between the magnets. Pig intestine from Yorkshire pigs was used to characterize the sensor. We are able to distinguish up to five distinct layers of the intestine from the large intestine. This sensing mechanism can indicate the operating surgeon the exact thickness of the tissue compressed between the two magnets. The surgeon can thus be sure of formation of a clean anastomosis and avoid the likelihood of the magnets sliding away or uncoupling.
dc.publisherThe American Society of Mechanical Engineers (ASME)
dc.titleInductive Sensing to Detect Tissue Thickness Between Magnets for Potential Application in Magnetic Compression Based Anastomosis
dc.typeJournal Paper
dc.journal.volume10
dc.journal.issue1
dc.journal.titleJournal of Medical Devices
dc.identifier.doi10.1115/1.4032615
dc.journal.fristpage11008
dc.journal.lastpage11008
dc.identifier.eissn1932-619X
dc.treeJournal of Medical Devices:;2016:;volume( 010 ):;issue: 001
dc.contenttypeFulltext


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record