Show simple item record

dc.contributor.authorLi, Huizhong
dc.contributor.authorAgarwal, Jay
dc.contributor.authorCoats, Brittany
dc.contributor.authorGale, Bruce K.
dc.date.accessioned2017-05-09T01:31:40Z
dc.date.available2017-05-09T01:31:40Z
dc.date.issued2016
dc.identifier.issn1932-6181
dc.identifier.othermed_010_01_011003.pdf
dc.identifier.urihttp://yetl.yabesh.ir/yetl/handle/yetl/161981
dc.description.abstractCurrently, endtoend anastomosis of blood vessels is performed using suturing, which is time consuming, expensive, and subject to large degrees of human error. One promising alternative is a ring–pin coupling device. This device has been shown to be useful for venous anastomosis, but lacks the versatility necessary for arterial applications. The purpose of this study was to optimize a vascular coupling design that could be used for arteries and veins of various sizes. To achieve this, finiteelement (FE) analysis was used to simulate the vessel–device interaction during anastomosis. Parametric simulations were performed to optimize the number of pins, the wing pivot point, and the pin offset of the design. The interaction of the coupler with various blood vessel sizes was also evaluated. Maximum strain in the vessel wall increased with the number of pins. The positions of the wings and pins were also important in dictating maximum strain, and improper dimensions lead to failure of the installation process. Extra force applied to the distal end of the vessel, or a supplementary tool, will be required during the coupler installation process to prevent vessels less than 3 mm inner diameter (0.5 mm wall thickness) from slipping off the coupler.
dc.publisherThe American Society of Mechanical Engineers (ASME)
dc.titleOptimization and Evaluation of a Vascular Coupling Device for End to End Anastomosis: A Finite Element Analysis
dc.typeJournal Paper
dc.journal.volume10
dc.journal.issue1
dc.journal.titleJournal of Medical Devices
dc.identifier.doi10.1115/1.4031810
dc.journal.fristpage11003
dc.journal.lastpage11003
dc.identifier.eissn1932-619X
dc.treeJournal of Medical Devices:;2016:;volume( 010 ):;issue: 001
dc.contenttypeFulltext


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record