Show simple item record

dc.contributor.authorTree, Michael
dc.contributor.authorWhite, Jason
dc.contributor.authorMidha, Prem
dc.contributor.authorKiblinger, Samantha
dc.contributor.authorYoganathan, Ajit
dc.date.accessioned2017-05-09T01:31:39Z
dc.date.available2017-05-09T01:31:39Z
dc.date.issued2016
dc.identifier.issn1932-6181
dc.identifier.othermed_010_01_011001.pdf
dc.identifier.urihttp://yetl.yabesh.ir/yetl/handle/yetl/161979
dc.description.abstractThe CardioMEMS heart failure (HF) system was tested for cardiac output (CO) measurement accuracy using an in vitro mock circulatory system. A software algorithm calculates CO based on analysis of the pressure waveform as measured from the pulmonary artery, where the CardioMEMS system resides. Calculated CO was compared to that from reference flow probe in the circulatory system model. CO measurements were compared over a clinically relevant range of stroke volumes and heart rates with normal, pulmonary hypertension (PH), decompensated left heart failure (DLHF), and combined DHLF + PH hemodynamic conditions. The CardioMEMS CO exhibited minimal fixed and proportional bias.
dc.publisherThe American Society of Mechanical Engineers (ASME)
dc.titleValidation of Cardiac Output as Reported by a Permanently Implanted Wireless Sensor
dc.typeJournal Paper
dc.journal.volume10
dc.journal.issue1
dc.journal.titleJournal of Medical Devices
dc.identifier.doi10.1115/1.4031799
dc.journal.fristpage11001
dc.journal.lastpage11001
dc.identifier.eissn1932-619X
dc.treeJournal of Medical Devices:;2016:;volume( 010 ):;issue: 001
dc.contenttypeFulltext


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record