Show simple item record

dc.contributor.authorKimura, Tadashi
dc.contributor.authorKubota, Makoto
dc.contributor.authorTaguchi, Tetsuya
dc.contributor.authorSuzuki, Naoki
dc.contributor.authorHattori, Asaki
dc.contributor.authorMarumo, Keishi
dc.date.accessioned2017-05-09T01:21:58Z
dc.date.available2017-05-09T01:21:58Z
dc.date.issued2015
dc.identifier.issn1932-6181
dc.identifier.othermed_009_04_044506.pdf
dc.identifier.urihttp://yetl.yabesh.ir/yetl/handle/yetl/159196
dc.description.abstractTo describe a loading device, we created for use with current computed tomography (CT) scanners and to evaluate its ability to reproduce loading conditions. Ten feet of five healthy volunteers (three men and two women) with no history of foot or ankle disorders and no foot pain were studied. Subjects lay on the device in the supine position with the legs extended and ankles in a neutral position. An axial load equivalent to their body weight was applied to the soles. A foot scan measured plantar contact area, plantar pressure, and the center of pressure under standing and weightbearing conditions. Differences between measurements were assessed using the paired t test with a twosided significance of 5%. No subjects complained of pain or discomfort during loading. Weightbearing measurements varied from standing measurements as follows: plantar contact area, −4.15 cm2; contact % (forefoot/midfoot/hindfoot), −0.61%/−0.17%/0.79%; plantar pressure, −1.47 N/cm2/−0.49 N/cm2/3.7 N/cm2 (6.40 N/cm2 overall); and center of pressure location (anterior–posterior/medial–lateral), 0.88%/0.36%. With the numbers available, no significant difference could be detected for any tested items. We were able to produce load distributions and intensities resembling standing conditions. Thus, the standing condition can be well produced on CT images if our device is used. Furthermore, images can be compared with those of healthy individuals to yield useful information for elucidating the pathophysiology and pathology of foot disorders.
dc.publisherThe American Society of Mechanical Engineers (ASME)
dc.titleAbility of a Novel Foot and Ankle Loading Device to Reproduce Loading Conditions in the Standing Position During Computed Tomography
dc.typeJournal Paper
dc.journal.volume9
dc.journal.issue4
dc.journal.titleJournal of Medical Devices
dc.identifier.doi10.1115/1.4031100
dc.journal.fristpage44506
dc.journal.lastpage44506
dc.identifier.eissn1932-619X
dc.treeJournal of Medical Devices:;2015:;volume( 009 ):;issue: 004
dc.contenttypeFulltext