    A Comparison of the Defect Rate Between Original Equipment Manufacturer and Reprocessed Single Use Bipolar and Ultrasound Diathermy Devices

    Thumbnail
    View/Open
    med_009_04_044501.pdf (51.65Kb)
    Date
    2015
    Author
    Loftus, Terrence J.
    Metadata
    Show full item record
    Abstract
    Reprocessing has emerged as an attempt to control the cost of singleuse bipolar and ultrasound diathermy devices despite limited data on defect rates. This study compares the defect rates, as reported by surgical teams, between original equipment manufacturer (OEM) singleuse bipolar and ultrasound diathermy devices and reprocessed (RP) devices. Data were retrospectively collected on 3112 devices over a 7month period for two types of bipolar and ultrasound diathermy devices. There is a significant difference (p < 0.001) in reported bipolar and ultrasound diathermy device defects between OEM and RP. OEM singleuse bipolar and ultrasound diathermy devices were reported to be defective more frequently than RP devices based on reports from the surgical team.
    URI
    http://yetl.yabesh.ir/yetl/handle/yetl/159191
    Collections
    • Journal of Medical Devices

    DSpace software copyright © 2002-2015  DuraSpace
    Contact Us | Send Feedback
    Theme by 
    @mire NV
     

     

    Browse

    All of DSpaceCommunities & CollectionsBy Issue DateAuthorsTitlesSubjectsThis CollectionBy Issue DateAuthorsTitlesSubjects

    My Account

    LoginRegister

    Statistics

    View Usage Statistics

    DSpace software copyright © 2002-2015  DuraSpace
    Contact Us | Send Feedback
    Theme by 
    @mire NV