Show simple item record

dc.contributor.authorZhao, Baoliang
dc.contributor.authorNelson, Carl A.
dc.date.accessioned2017-05-09T01:21:56Z
dc.date.available2017-05-09T01:21:56Z
dc.date.issued2015
dc.identifier.issn1932-6181
dc.identifier.othermed_009_04_041012.pdf
dc.identifier.urihttp://yetl.yabesh.ir/yetl/handle/yetl/159189
dc.description.abstractRobotic minimally invasive surgery (RMIS) has achieved success in various procedures; however, the lack of haptic feedback is considered by some to be a limiting factor. The typical method to acquire tool–tissue reaction forces is attaching force sensors on surgical tools, but this complicates sterilization and makes the tool bulky. This paper explores the feasibility of using motor current to estimate tooltissue forces and demonstrates acceptable results in terms of time delay and accuracy. This sensorless force estimation method sheds new light on the possibility of equipping existing robotic surgical systems with haptic interfaces that require no sensors and are compatible with existing sterilization methods.
dc.publisherThe American Society of Mechanical Engineers (ASME)
dc.titleSensorless Force Sensing for Minimally Invasive Surgery
dc.typeJournal Paper
dc.journal.volume9
dc.journal.issue4
dc.journal.titleJournal of Medical Devices
dc.identifier.doi10.1115/1.4031282
dc.journal.fristpage41012
dc.journal.lastpage41012
dc.identifier.eissn1932-619X
dc.treeJournal of Medical Devices:;2015:;volume( 009 ):;issue: 004
dc.contenttypeFulltext


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record