Show simple item record

dc.contributor.authorNelson, Carl A.
dc.contributor.authorStolle, Cale J.
dc.contributor.authorBurnfield, Judith M.
dc.contributor.authorBuster, Thad W.
dc.date.accessioned2017-05-09T01:21:56Z
dc.date.available2017-05-09T01:21:56Z
dc.date.issued2015
dc.identifier.issn1932-6181
dc.identifier.othermed_009_04_041010.pdf
dc.identifier.urihttp://yetl.yabesh.ir/yetl/handle/yetl/159187
dc.description.abstractEffective gait therapy is critical to children who have difficulty walking due to developmental, neurologic, or orthopedic conditions. Current gait training technologies can be cost prohibitive and often do not address the needs of children of varying sizes. In addition, clinicians often need to provide significant physical assistance to children with profound weakness. Based on the success of an ellipticalbased adultsized intelligently controlled assistive rehabilitation elliptical (ICARE) system for gait training, a modified technology was proposed to address the needs of younger/smaller children. The new design relied on a screwandslider joint to adjust the effective length of the crank link in the elliptical mechanism, reducing the step length and stride height simultaneously. The new trajectories of the foot pedal were normalized against stride length and showed nearly identical trajectories between pediatric strides and adult strides. Simulation results and human usability studies verified that the design was feasible.
dc.publisherThe American Society of Mechanical Engineers (ASME)
dc.titleModification of the ICARE System for Pediatric Therapy
dc.typeJournal Paper
dc.journal.volume9
dc.journal.issue4
dc.journal.titleJournal of Medical Devices
dc.identifier.doi10.1115/1.4030276
dc.journal.fristpage41010
dc.journal.lastpage41010
dc.identifier.eissn1932-619X
dc.treeJournal of Medical Devices:;2015:;volume( 009 ):;issue: 004
dc.contenttypeFulltext