Show simple item record

dc.contributor.authorAskari, Ehsan
dc.contributor.authorFlores, Paulo
dc.contributor.authorDabirrahmani, Danأ¨
dc.contributor.authorAppleyard, Richard
dc.date.accessioned2017-05-09T01:15:39Z
dc.date.available2017-05-09T01:15:39Z
dc.date.issued2015
dc.identifier.issn1555-1415
dc.identifier.othercnd_010_02_024502.pdf
dc.identifier.urihttp://yetl.yabesh.ir/yetl/handle/yetl/157273
dc.description.abstractA ceramiconceramic (CoC) hip prosthesis with clearance is modeled as a multibody dynamics system for the purpose of studying hip squeaking. A continuous contact force model provides the intrajoint forces developed at the hip joint. Friction effects due to the relative motion are also considered. A FFT analysis of the audible sounds from CoC hip acceleration is carried out to analyze hip squeaking. The effects of friction, hip implant size, and the head initial position on hip squeaking and the trajectory of femoral head are analyzed and discussed. It was shown that the causes of hip squeaking are stick/slip, frictioninduced vibration, and the femoral head angular speed and force changes.
dc.publisherThe American Society of Mechanical Engineers (ASME)
dc.titleA Computational Analysis of Squeaking Hip Prostheses
dc.typeJournal Paper
dc.journal.volume10
dc.journal.issue2
dc.journal.titleJournal of Computational and Nonlinear Dynamics
dc.identifier.doi10.1115/1.4028109
dc.journal.fristpage24502
dc.journal.lastpage24502
dc.identifier.eissn1555-1423
dc.treeJournal of Computational and Nonlinear Dynamics:;2015:;volume( 010 ):;issue: 002
dc.contenttypeFulltext


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record