dc.description.abstractMagnetic resonance (MR) thermometry is a valuable method for characterizing thermal fields generated by high intensity focused ultrasound (HIFU) transducers in tissue phantoms and excised tissues. However, infiltration of noise signals generated by external rf sources into the scanner orifice limits the ability of the scanner to measure temperature rise during the heating or ablation phase. In this study, magnetic resonance interferometry (MRI) monitored HIFU ablations are performed on freshly excised porcine liver samples, at varying sonication times, 20 s, 30 s, and 40 s at a constant acoustic intensity level of 1244 W/cm2. Temperature throughout the procedure was measured using proton resonant frequency MR thermometry. Without filtering, reliable temperature measurements during the heating phase could not be obtained since temperature maps appeared blurred and analysis was impossible. Also, measurements acquired during the cooling phase decayed manifested an unrealistically slow rate of temperature decay. This abnormally slow rate was confirmed with computational results. A low-pass RC filter circuit was subsequently incorporated into the experimental setup to prevent infiltration of noise signals in the MRI orifice. This modified RC filter circuit allowed noninvasive measurement of the HIFU induced temperature rise during the heating phase followed by temperature decay during cooling. The measured data were within 13% agreement with the temperature rise computed by solving the acoustic and heat equations.
dc.publisherThe American Society of Mechanical Engineers (ASME)
dc.titleReduction of Noise From MR Thermometry Measurements During HIFU Characterization Procedures
dc.typeJournal Paper
dc.journal.volume2
dc.journal.issue2
dc.journal.titleJournal of Nanotechnology in Engineering and Medicine
dc.identifier.doi10.1115/1.4003861
dc.journal.fristpage24501
dc.identifier.eissn1949-2952
dc.keywordsTemperature
dc.keywordsCooling
dc.keywordsFiltration
dc.keywordsTemperature measurement
dc.keywordsNoise (Sound)
dc.keywordsAblation (Vaporization technology)
dc.keywordsMagnetic resonance imaging
dc.keywordsHeating
dc.keywordsBiological tissues
dc.keywordsTransducers
dc.keywordsCircuits
dc.keywordsMagnetic resonance
dc.keywordsLiver
dc.keywordsAcoustics
dc.keywordsSignals
dc.keywordsFilters
dc.keywordsPhantoms
dc.keywordsAcoustic intensity AND Interferometry
dc.treeJournal of Nanotechnology in Engineering and Medicine:;2011:;volume( 002 ):;issue: 002
dc.contenttypeFulltext


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record