Show simple item record

dc.contributor.authorJustin M. Stinnett-Donnelly
dc.contributor.authorJeffrey J. MacLean
dc.contributor.authorJames C. Iatridis
dc.date.accessioned2017-05-09T00:25:19Z
dc.date.available2017-05-09T00:25:19Z
dc.date.copyrightMarch, 2007
dc.date.issued2007
dc.identifier.issn1932-6181
dc.identifier.otherJMDOA4-27980#56_1.pdf
dc.identifier.urihttp://yetl.yabesh.ir/yetl/handle/yetl/136597
dc.description.abstractThe rat tail intervertebral disc has emerged as an important model to examine the mechanisms for mechanically induced degeneration and remodeling. Previous methods used to apply high precision axial compressive loading to a rat tail intervertebral disc in vivo either required anesthesia, or the permanent mounting of a loading device to the animal, and were not well described in the literature. Therefore, a new device to apply compressive loading to the rat tail intervertebral disc was developed and validated. The rat tail compressive loading system utilized a pneumatically driven device weighing 18g, and was capable of delivering a 12.6N sinusoidal or square waveform at frequencies up to 1.0Hz. The system improved on previous methods in its modular construction, relative ease of fabrication, compatibility with existing tail model technology and overall cost effectiveness. The removable system eliminated the need for anesthesia and through a modular, cost effective, design allowed for the simultaneous loading of multiple animals. This system expanded the ability to accurately, ethically, and efficiently apply dynamic compressive loads to the rat tail intervertebral disc for extended periods of time in order to address questions related to disc mechanobiology.
dc.publisherThe American Society of Mechanical Engineers (ASME)
dc.titleA Removable Precision Device for In-Vivo Mechanical Compression of Rat Tail Intervertebral Discs
dc.typeJournal Paper
dc.journal.volume1
dc.journal.issue1
dc.journal.titleJournal of Medical Devices
dc.identifier.doi10.1115/1.2355692
dc.journal.fristpage56
dc.journal.lastpage61
dc.identifier.eissn1932-619X
dc.keywordsForce
dc.keywordsStress
dc.keywordsDesign
dc.keywordsDisks
dc.keywordsAccuracy
dc.keywordsCompression
dc.keywordsIntervertebral discs
dc.keywordsSurgery
dc.keywordsPneumatic systems
dc.keywordsControl systems AND Frequency
dc.treeJournal of Medical Devices:;2007:;volume( 001 ):;issue: 001
dc.contenttypeFulltext


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record