Show simple item record

dc.contributor.authorD. Croft
dc.contributor.authorG. Shed
dc.contributor.authorS. Devasia
dc.date.accessioned2017-05-09T00:04:31Z
dc.date.available2017-05-09T00:04:31Z
dc.date.copyrightMarch, 2001
dc.date.issued2001
dc.identifier.issn0022-0434
dc.identifier.otherJDSMAA-26279#35_1.pdf
dc.identifier.urihttp://yetl.yabesh.ir/yetl/handle/yetl/124988
dc.description.abstractThis article studies ultra-high-precision positioning with piezoactuators and illustrates the results with an example Scanning Probe Microscopy (SPM) application. Loss of positioning precision in piezoactuators occurs (1) due to hysteresis during long range applications, (2) due to creep effects when positioning is needed over extended periods of time, and (3) due to induced vibrations during high-speed positioning. This loss in precision restricts the use of piezoactuators in high-speed positioning applications like SPM-based nanofabrication, and ultra-high-precision optical systems. An integrated inversion-based approach is presented in this article to compensate for all three adverse affects—creep, hysteresis, and vibrations. The method is applied to an Atomic Force Microscope (AFM) and experimental results are presented that demonstrate substantial improvements in positioning precision and operating speed.
dc.publisherThe American Society of Mechanical Engineers (ASME)
dc.titleCreep, Hysteresis, and Vibration Compensation for Piezoactuators: Atomic Force Microscopy Application
dc.typeJournal Paper
dc.journal.volume123
dc.journal.issue1
dc.journal.titleJournal of Dynamic Systems, Measurement, and Control
dc.identifier.doi10.1115/1.1341197
dc.journal.fristpage35
dc.journal.lastpage43
dc.identifier.eissn1528-9028
dc.treeJournal of Dynamic Systems, Measurement, and Control:;2001:;volume( 123 ):;issue: 001
dc.contenttypeFulltext


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record