Show simple item record

contributor authorKumar, Vijay
contributor authorJoseph, Antony
contributor authorPrabhudesai, R. G.
contributor authorPrabhudesai, S.
contributor authorNagvekar, Surekha
contributor authorDamodaran, Vimala
date accessioned2017-06-09T17:23:00Z
date available2017-06-09T17:23:00Z
date copyright2005/12/01
date issued2005
identifier issn0739-0572
identifier otherams-84196.pdf
identifier urihttp://onlinelibrary.yabesh.ir/handle/yetl/4227505
description abstractSimultaneous calibrations of three temperature-compensated piezoresistive ruggedized precision ?absolute? pressure transducers (Honeywell model PPTR0040AP5VB-BD), which have been designed specially for long-term coastal oceanographic and limnological measurements, have been carried out at four differing temperatures (10°, 20°, 30°, and 40°C) to evaluate their suitability for such applications. The full-scale pressure range of these shallow water absolute pressure sensors is ≈ 2800 hPa (equivalent to water depth of ≈ 18 m). Measurement results have been used to examine the transducers? performance indicators, such as zero-point offset, accuracy, linearity, hysteresis, temperature sensitivity, and slope. Differing piezoresistive ruggedized precision absolute pressure transducers (PPTRs) exhibited differing zero-point offset values, ranging from 2 to ?79 hPa. Temperature sensitivity of zero-point offset was ≈0.3 hPa over the temperature range 10°?40°C. The mean hysteresis over the full-scale absolute pressure range (≈2800 hPa) varied from approximately 2 to 8 hPa over the temperature range 10°?40°C. The slope of the least squares?fitted linear graph (taking the mean of ascending and descending pressures) was close to the ideal value of unity (deviation from 1 over the temperature range 10°?40°C was in the range of ?0.001 to +0.005). Linearity was excellent, its mean over the entire pressure range being between ≈ ?0.006% and 0.008% of full-scale (FS) over the above temperature range. The worst performance was exhibited at input pressures below ≈1500 hPa. Zero-point offset has played a significant role in deteriorating the accuracy of the PPTR, the mean accuracy (within ≈0.1% and ?5%) having been exhibited by those transducers having offsets of 2 and ?79 hPa, respectively. The mean accuracy exhibited temperature sensitivity of ≈1% in the range 10°?20°C and negligible sensitivity beyond 20°C. Use of a calibration equation significantly improved the mean static accuracy obtainable from the PPTR, to between ?0.04% and 0.01% of FS. Evaluation results have indicated that a suitably calibrated temperature-compensated Honeywell PPTR provides an alternate cost-effective means for pressure measurements for coastal oceanographic and limnological studies.
publisherAmerican Meteorological Society
titlePerformance Evaluation of Honeywell Silicon Piezoresistive Pressure Transducers for Oceanographic and Limnological Measurements
typeJournal Paper
journal volume22
journal issue12
journal titleJournal of Atmospheric and Oceanic Technology
identifier doi10.1175/JTECH1812.1
journal fristpage1933
journal lastpage1939
treeJournal of Atmospheric and Oceanic Technology:;2005:;volume( 022 ):;issue: 012
contenttypeFulltext