Show simple item record

contributor authorChavez, Francisco P.
contributor authorWright, Dave
contributor authorHerlien, Robert
contributor authorKelley, Michael
contributor authorShane, Farley
contributor authorStrutton, Peter G.
date accessioned2017-06-09T14:17:58Z
date available2017-06-09T14:17:58Z
date copyright2000/02/01
date issued2000
identifier issn0739-0572
identifier otherams-1675.pdf
identifier urihttp://onlinelibrary.yabesh.ir/handle/yetl/4152567
description abstractA shutter mechanism for reducing the effects of biofouling on bio-optical instruments deployed on oceanographic moorings has been designed, built, and tested. The initial development was carried out on a spectroradiometer. The optics of the spectroradiometer are protected by copper shutters that rotate out of the field of view prior to a measurement and rotate back after the measurement is completed. The shutter system can sense an obstruction and, if one is detected, attempt to rotate in the opposite direction. The controlling software stores the home position in the memory so the shutter can return to cover the optics, irrespective of direction of rotation. The system has been tested in the equatorial Pacific, where it has provided five months of data that are unaffected by biofouling.
publisherAmerican Meteorological Society
titleA Device for Protecting Moored Spectroradiometers from Biofouling
typeJournal Paper
journal volume17
journal issue2
journal titleJournal of Atmospheric and Oceanic Technology
identifier doi10.1175/1520-0426(2000)017<0215:ADFPMS>2.0.CO;2
journal fristpage215
journal lastpage219
treeJournal of Atmospheric and Oceanic Technology:;2000:;volume( 017 ):;issue: 002
contenttypeFulltext


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record